The FC300B is a glass combination ion selective electrode (ISE) for the determination of sodium (Na+) ions in solution.
The selective glass membrane produces a potential change due to the exchange of sodium ions on the surface of the membrane and the sample. The internal sensing elements are housed within a sturdy glass body.
The FC300B is ideal for a variety of applications in laboratories, food and beverage production, and water quality analysis.
The FC300B sodium ion selective electrode is a combination electrode that has a silver/silver chloride (Ag/AgCl) reference and a sodium ion selective electrode. Both are housed within a single glass body design that uses an ammonium chloride/silver chloride electrolyte for a fill solution.
The sodium sensitive membrane should be conditioned before performing a calibration. The conditioning of the electrode is a two-step process in which the glass is etched then rehydrated. The first step is etching. This process removes a thin layer of sensitive glass from the sodium membrane, exposing a sodium-containing fresh surface. The second step is rehydration in which the probe is placed in storage solution to rehydrate the glass membrane. Allowing the glass to rehydrate after etching provides optimum conditions for the exchange of sodium ions between the membrane and solution.
For the FC300B to measure sodium accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the sodium ionic strength adjuster (ISA) solution. The sodium ISA is important for two reasons. One is that the ISA buffers the pH of the standard and sample to above pH 9.8. This removes any interference from hydrogen ions that occur at lower pH’s. The second purpose of the ISA is to ensure that the activity coefficient is constant so the concentration of free sodium ions can be measured. Having solutions with a constant ionic strength reduces the margin of error between measurements of different samples at different concentrations.
The FC300B sodium glass ISE can measure from 10µM (0.23 mg/L) to 1M (22,990 mg/L).
Multiply |
|
moles/L (M) Na+ to ppm (mg/L) Na+ |
22990 |
ppm (mg/L) Na+ to M (moles/L) Na+ |
4.35 x 10-5 |
ppm (mg/L) Na+ to ppm (mg/L) NaCl |
2.54197 |
ppm (mg/L) NaCl to ppm (mg/L) Na+ |
0.39339 |
Spherical Glass Tip - The spherical shaped tip design allows for a wide area of contact with the sample, permitting a faster electrode response with a higher degree of stability.
Glass Electrode Body - The glass electrode body is suitable for a wide range of applications due to its chemical resistance. The glass electrode is compatible with many non-aqueous solvents and other aggressive chemicals. Glass is also resistant to many forms of radiation, such as ultraviolet radiation.
BNC Connection - The FC300B has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Type | glass; combination |
---|---|
Measurement Range | Sodium (Na+) 1.0M to 1 x 10-5M, 22,990 to 0.23 mg/L (ppm) |
Optimum pH Range | pH 9.75 to 14 |
Temperature Range | 0 to 80 °C |
Approximate Slope | +57 |
Diameter | 12 mm |
Overall Length | 120 mm |
Body Material | glass |
Cable | coaxial; 1 m (3.3’) |
Connection | BNC |