Digital sensor to measure conductive conductivity especially in pure media, for operation on TriBox controllers and HS100 DIN G2 rail module. The digital technology ensures secure and interference-free signal transmission from the sensor to the controller.
Specification
Measurement technology
|
Conductivity
|
|
Measurement principle
|
Conductivity with two graphite electrodes
|
|
|
|
|
Parameter
|
Conductivity
|
|
Measurement range
|
0.00 µS… 20000 µS
|
|
Measurement accuracy
|
±0.5 µS at 20 µS
± 5 µS at 200 µS
± 50 µS at 2000 µS
± 500 µS at 20000 µS
|
|
Response time
|
T90 < 60s
|
|
Temperature compensation
|
Via NTC
|
|
|
|
|
Housing material
|
PVC housing, graphite electrodes
|
|
Dimensions (L x Ø)
|
220 mm x 33 mm
|
~ 8.7“ x 1.3“
|
|
|
|
Interface
|
RS485 (Modbus RTU)
|
|
Power supply
|
12…24 VDC
|
|
Connection
|
8-pin M12 connector, cable length 2 m or 10 m
|
|
Maintenance interval
|
2 years
|
|
System compatibility
|
Modbus RTU
|
|
Warranty
|
1 year (EU & US: 2 years) on electronics;
wear parts are excluded from the warranty
|
|
|
|
|
Process pressure
|
10 bar
|
~ 145 psig
|
Calibration method
|
One-point calibration with standard measuring solution
|
|
|
|
|
Process temperature
|
0…50°C
|
~ +32 °F to +122 °F
|