The TTurb is a digital sensor used for optical turbidity measurement according to the 90° IR scattered light method. Depending on the sensor version, it can be used in both pure water up to 100 FNU as well as in raw, waste and process waters up to 1000 FNU. The TTurb is available with a 10-meter or a 2-meter cable.
Specification
Light Source
|
LED
|
||
Detector
|
Photo diode
|
||
Measurement principle
|
Nephelometry
|
||
|
|||
Parameter
|
Turbidity as FNU; mg/L; NTU; TSSeq
|
||
Measurement range
|
0...100, 0...400, 0...1000 FNU
|
||
Measurement accuracy
|
± (5 % + 0.5)
|
||
Detection limit
|
0.5 FNU for TTurb 100
2 FNU for TTurb 400
2 FNU for TTurb 1000
|
||
Measurement wavelength
|
860 nm, FWHM 30 nm
|
||
T100 response time
|
6 s
|
||
Measurement interval
|
≥ 3 s
|
||
|
|||
Housing material
|
PET / POM / NBR
|
||
Dimensions (L x Ø)
|
170 x 36 mm
|
~ 6.7 x 1.4
|
|
Weight
|
0.3 kg
|
~ 0.7 lbs
|
|
|
|||
Interface
|
Ethernet (TCP/IP)
RS-485 (Modbus RTU)
|
||
Power consumption
|
typically <0.9 W
with network < 1.5 W
|
||
Power supply
|
12...24 VDC (± 10 %)
|
||
Connector
|
8-pin M12-plug
|
||
|
|||
Maintenance effort
|
≤ 0,5 h/Monat (typisch)
|
||
Calibration/maintenance interval
|
24 Monate
|
||
System compatibility
|
Modbus RTU
|
||
Warranty
|
1 Year (EU & US: 2 years) on electronics; All wearing parts are not included in the warranty
|
||
|
|||
Max. pressure with fixed cable
|
3 bar
|
~ 43.5 psig
|
|
Max. pressure in flow cell
|
1 bar, 2...4 L/min
|
~ 14.5 psig, 0.5 to 1 gpm
|
|
Protection type
|
IP68
|
NEMA 6P
|
|
|
|||
Sample temperature
|
0...+40 °C
|
~ +32 °F... +104 °F
|
|
Ambient temperature
|
0...+40 °C
|
~ +32 °F... +104 °F
|
|
Storage temperature
|
0...+80 °C
|
~ +32 °F... +176 °F
|
|
Inflow velocity
|
maximum 0.1 m/second
|
maximum ~ 0.33 fps
|
TTurb100
|
0...100 NTU
|
TTurb400
|
0...400 NTU
|
TTurb1000
|
0...1000 NTU
|