Genie E EDI Pure Water System produces EDI water (Type II water) from tap water directly. Resistivity of the product water is above 5 MΩ•cm at 25 ℃ , which meets or exceeds Type II water quality as defined by ASTM, CAP, CLSI and ISO 3696 / BS 3997 and also complies with the purified water requirements from the European and U.S. Pharmacopoeia.
The system is CE/RoHS certified. RephiLe provides full documents support to meet user’s GMP, GSP, GAP, GCP, GLP requirements.
Design & Functions
Effective TOC level controlment
Easy Operation and Maintenance
Cartridges and Parts
1. EDI module
Internal EDI module removes the remaining ions, further reduce the content of organic compounds and has a certain capacity to prevent the growth of bacteria from RO water.
Ion exchange resins are continuously regenerated by an electric field. No hazardous chemical regeneration or costly resin replacement is needed.
EDI module has a high degree of automation and minimal electricity consumption by using the equivalent of the energy.
2. Cartridge
Internal P Pack cartridge removes oxidants, organics, particles and scaling ions to prevent them from fouling the RO membranes and the EDI module.
AC Pack contains special activated carbon to remove organic impurities and free chlorines.
3. Control console
8-inch console controls system and peripheral devices (Such as leak protector and tank sanitization module). All operations can be done on the console by touching the screen.
The control console can be placed on the bench or inside a drawer for further flexibility and space saving.
The console screen and dispenser screens are water-drop proof. You can operate the console and dispenser with latex gloves on.
The system has two level password protection on display menu for safety purposes.
4. Dispenser
The volumetric function can automatically deliver the water volume you need once it is set up.
With the adjusting button on a control console or a dispenser, you can dispense water at the flow rate you need up to 2 liters per minute.
A dispenser handle can be set on the dispenser stand, on the main system, or even hanging onto other places to free up maximum bench space.
5. Final Filter
Various final filters (optional) are applicable to ensure ultrapure water without particles, bacteria or pyrogen.
6. Level sensor
The tank integrated continuous level sensor measures water level within the tank and manages the system to start or stop producing water automatically based on requirements.
Water level and quality inside the tank can be viewed from the control console.
EDI pure water can be used in many areas. Here are some typical applications.
Feed Water Requirements | |
Tap water | Municipal water, Conductivity < 2000 μS/cm (1000 ppm) |
Operating temperature | 5 - 35 ℃ |
Feed water pressure | 1 - 6 bar (15 - 90 psi) |
Water Flow Rate | |
Water production rate (@ 25℃) | 5, 10, 15 L/hr |
Water dispensing rate | Up to 2.0 L/min |
Product Water Quality | |
RO rejection | 97 to 98% ionic rejection (new RO cartridge) > 99% organic rejection > 99% particulates & bacteria rejection |
Water resistivity (@ 25℃) | > 5 MΩ·cm (typically 10 - 16 MΩ·cm) |
TOC* | < 30 ppb |
Product Water Quality Stored in Tank (for tank recirculation models) | |
Resistivity (@25℃) | > 1 MΩ·cm (default), adjustable |
Dimensions | |
Main system dimensions Width × depth × height | 32 × 44 × 54 cm (12.6 × 17.3 × 21.3 in) |
Dispenser dimensions Width × depth × height | 21 × 29 × 61 cm (8.3 × 11.4 × 24.0 in) |
Weight | 20 kg |
Input Voltage | 110 - 240 VAC |
Operating Voltage | 24 VDC |
Main system power | < 200 W |
*Product water quality may vary due to local feed water conditions.