Description
Salt-bath furnace TS ../50 is designed especially for chemical tempering of glass in the laboratory. Chemical tempering is a process used to strengthen thin glass with wall thicknesses of just a few millimeters. The benefit of chemical pre-stressing is that the surface quality remains intact. Almost all glass with a high sodium content can be strengthened through ion exchange.
- Tmax 500 °C
- Salt-bath furnace in compact design with salt-bath and pre-heated-/cooling chamber above the salt-bath
- Exclusive use of insulation materials without categorization according to EC Regulation No 1272/2008 (CLP). This explicitly means that alumino silicate wool, also known as “refractory ceramic fiber” (RCF), which is classified and possibly carcinogenic, is not used.
- Bath temperature control
- Indirect heating of the preheated chamber from the salt-bath
- Automatic, time controlled movement from the preheating chamber into the salt-bath and back
- Electrical door lock
- Crucible made of hiqh-quality CrNi steel
- Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
- Defined application within the constraints of the operating instructions
- NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
- Controller P570 (50 programs with each 40 segments)
Additional Equipment
- Hood for connection to local exhaust system
- Charging basket according to customers drawing
- Active heating for the preheated chamber
- PLC
- Process control and documentation via Nabertherm Control Center (NCC) for monitoring, documentation and control
Model |
Tmax |
Inner dimensions crucible in mm |
Volume |
Outer dimensions2 in mm |
Heating power |
Electrical |
Weight |
|
°C1 |
w |
d |
h |
in l |
W |
D |
H |
in kW3 |
connection* |
in kg |
TS 8/50 |
500 |
300 |
100 |
100 |
8 |
1600 |
1050 |
2400 |
2.0 |
3-phase |
650 |
TS 90/50 |
500 |
650 |
300 |
450 |
90 |
1600 |
1050 |
2400 |
20.0 |
3-phase |
700 |
1Salt-bath temperature
2External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
3Depending on furnace design connected load might be higher |