General specifications |
Measuring functions |
|
3-pole earth ground resistance, 2-pole ac resistance of a conductor, Interference voltage |
|
Intrinsic error |
|
Refers to the reference temperature range and is guaranteed for one year |
|
Measuring rate |
|
Battery[1] |
|
One 9 volt alkaline (LR61) |
|
Battery condition |
|
LO-BAT is displayed if voltage drops below 6.5 V |
|
Voltages |
Between jacks H/C2 and E/C1: |
250 Veff maximum (effective voltage) |
Between jacks S/P2 and E/C1: |
250 Veff maximum |
|
Climatic class |
|
VDE/VDI 3540 RZ (conforming to KWG as per DIN 40040, 4/87) |
|
Temperature performance[2] |
Working: |
-10 °C to +50 °C (+14 °F to +122 °F) |
Operating: |
0 °C to +35 °C (+32 °F to +95 °F) |
Storage: |
-20 °C to +60 °C (+68 °F to +140 °F) |
Reference: |
+23 °C ± 2 °C (+73 °F ± 4 °F) |
|
Temperature coefficient |
|
± 0.1 % of range per degree Kelvin |
|
Safety |
|
IEC/EN 61010-1, 600 V CAT II, pollution degree 2 |
|
Dimensions |
|
113 mm x 54 mm x 216 mm (4.5 in x 2.1 in x 8.5 in), including holster |
|
Weight |
|
850 g (1.9 lb), including standard accessories, volume approximately 600 cm3 |
|
Note: |
|
[1]If the tester is not going to be used, or is being stored for a long period, remove the battery and store separately from the tester to avoid damage from battery leakage. |
|
Note: |
|
[2]The four temperature ranges for the tester exists to satisfy European Standards requirements; the instrument can be used over the full working temperature range by using the temperature coefficient to calculate accuracy at the ambient temperature of use. |
|