详情
PCT200i is specially designed for current transformer characteristic test. PCT200i owns more test items and higher technical parameters.This CT analyzer can finish the measurement class CT and TYP class CT.The additional functions include ratio and angle differential tests (the measurement points in the GB standard),FS and inductance,steady-state parameters,peak error and transient parameters,etc.
Features
- Test CT according to IEC60044-1/6, IEC61869-2, C57.13
- Can test all types of CT, including bushing and GIS type CT
- One test can get all CT parameters
- Fast test time for excitation test
- 10kV knee point test: 50s
- 1kV knee point test: 15s
- Ratio check up to 30,000:1
- Knee point check up to 45,000V
- PT test functions are also in-built
- Test results can be saved to flash disk
- Optional Excel format report generation
- Large color LCD display
- Weight : 10/14kg
Main Specifications
<table width:100%;"="" style="width:100%;" cellspacing="0" cellpadding="5" border="1"> Voltage output 0-120V Current output 0~5Arms(15A peak) Ratio check Range 35,000:1 - 45,000:5 Accuracy 0.1% Knee point check max. 45,000V Burden measurement Range 0.001-1000 ohm Accuracy ±0.1%±1mΩ Gur Winding resistance test Range 0.001-1000 ohm Accuracy ±0.1%±1mΩ Gur Phase displacement check Accuracy 3 min Typ Resolution 0.1 min Other test functions Accuracy limiting factor Yes Instrument security factor Yes Remanence flux Yes Secondary time constant Yes Unsaturated inductance Yes Saturated inductance Yes Symmetrical short circuit factor Yes Accuracy limiting voltage Yes Transient dimensioning factor Yes Peak instantaneous error Yes Composite error Yes Dimensioins 470×200×245mm (W×H×L) Weight 10/14kg
更多 Labmetrik Electrical Sdn Bhd 相关资料