Key features
Specifications | |
Infrared temperature range | -40°C to 800°C (-40°F to 1472°F) |
Infrared accuracy |
< 0°C (32°F): ±(1.0°C (±2.0°F) + 0.1°/1°C or°F); > 0°C (32°F): ±1% or ±1.0°C (±2.0°F), whichever is greater |
Display resolution | 0.1°C / 0.1°F |
Infrared spectral response | 8 µm to 14 µm |
Infrared response time | < 500 ms |
Thermocouple type-k input temperature range | -270°C to 1372°C (-454°F to 2501°F) |
Thermocouple type-k input accuracy |
< -40°C: ±(1°C + 0.2°/1°C) ≥ -40°C: ±1% or 1°C, whichever is greater < -40°F: ±(2°F + 0.2°/1°F) ≥ -40°F: ±1% or 2°F, whichever is greater |
D:S (distance to measurement spot size) | 50:1 |
Laser sighting | Single-point laser |
Minimum spot size | 19 mm (0.75 in) |
Emissivity adjustment | By built-in table of common materials or digitally adjustable from 0.10 to 1.00 by 0.01 |
Data storage | 99 points |
Hi/Low alarms | Audible and two-color visual |
Min/Max/Avg/Dif | Yes |
Display | Dot matrix with function menus |
Backlight | Two levels, normal and extra bright for darker environments |
Trigger lock | Yes |
Switchable celsius and fahrenheit | Yes |
Power | 2 AAA/LR03 type-approved batteries (For a list of type-approved batteries, refer to Product Safety Instructions.) |
Battery life |
4 hours with laser and backlight on; 100 hours with laser and backlight off, at 100% duty cycle |
Operating temperature | 0°C to 50°C (32°F to 122°F) |
Storage temperature | -20°C to 60°C (-40°F to 140°F) |
Bead thermocouple type-k range | -40°C to 260°C (-40°F to 500°F) |
Bead thermocouple type-k accuracy |
±1.1°C (2.0°F) from 0°C to 260°C (32 °F to 500°F), typically within 1.1°C (2.0°F) from -40°C to 0°C (-40°F to 32°F) |