GL7000

GL7000

分类: DATA PLATFORM 当前有货
欲了解更多详情,请游览我们的官方网站 elsotech.com.my

详情

GL7000 specifications

Item Description
Model number GL7000
Number of module Attached to up to 10 modules (*1)
Number of input channels Max. 112 channels in 1 of GL7000
External Input/Output signals (*2) Input  Start/Stop, External trigger, External sampling, Auto balance (*3)
Signal type: Contact (relay), Open collector, Voltage
Output Trigger, Busy (*3), Alarm (10 channels) (*4)
Signal type: Open collector (pulled-up by resistor 10 kΩ)
Trigger, Alarm function Trigger action Start or stop capturing data by the trigger
Trigger repeat Enabled (ON): Automatically re-armed for the next data capture function
Disabled (OFF): Data capture is completed in a single trigger
Trigger source Start: Off, Measured signal, Alarm, External signal, Clock, Week or Time
Stop: Off, Measured signal, Alarm, External signal, Clock, Week or Time
Trigger determination conditions for measured signal Combination: OR or AND condition at the level of signal or edge of signal
Analog: Higher/Rising, Lower/Falling, Window-in, Window-out
Logic (*5): Higher/Rising, Lower/Falling
Pulse (*5): Higher/Rising, Lower/Falling, Window-in, Window-out
Alarm determination condition (*6) Combination: OR or AND condition at the level of signal or edge of signal
Analog: Higher/Rising, Lower/Falling, Window-in, Window-out
Logic (*5): Higher/Rising, Lower/Falling
Pulse (*5): Higher/Rising, Lower/Falling, Window-in, Window-out
Alarm output 10 channels
Pre-trigger (*7) Number of data before trigger: Up to specified number of captured data
Calculation function Between channels Addition, Subtraction, Multiplication and Division for two analog inputs (Sampling speed is limited up to 10 Samples/s (100 ms interval). Available arithmetic element and the output destination is the analog input channel 1 to 100.)
Statistical Select two calculations from Average, Peak, Max., Min. in real time and replay (*8)
Move function of the display range Beginning, center or end of the data, Trigger point, Specific time (absolute, relative), Call cursor
Search function Search for analog signal levels, logic signal pattern, pulse signal levels or alarm point in captured data
Annotation function Comment can be set in each channel (up to 31 alphanumeric characters)
Message / Marker Functions Message: The registered messages or entered message is able to be recorded for any timing. Up to 8 messages can be pre-registered.
Marker: Marker is able to record for occurring alarm or power failure.
Resume Resume automatically in the same condition after power is recovered as when the power failure occurred during data capture (*9)
FFT analysis function
(Firmware ver. 1.20 or later)
Analyzing frequency range 0.08, 0.2, 0.4, 0.8, 1.6, 2, 3.2, 4, 8, 20, 40, 80, 200, 400, 800 Hz, 2, 4, 8, 20, 40, 80, 200, 400 kHz
Number of points 500, 1000, 2000, 4000, 10000
Window function Rectangular, Hanning, Hamming, Blackman, Flat-top, Exponential
Averaging Summation average, Exponential average, Peak hold
Channels 4 channels
Functions Y-T, Linear, Power, PSD, Cross, Transfer function, Coherence, COP
Display mode Single display, Dual display, Nyquist
Interface to PC Ethernet (10 BASE-T/100 BASE-TX), USB 2.0 (High speed)
Network function WEB server, FTP server, FTP client, NTP client, DHCP client
USB drive mode Emulate the USB memory device (*10)
Storage device Built-in RAM (2 million samples, built-in amplifier module),
Flash memory (2 GB, built-in the main module)
External (*11) SD card (Support SDHC, up to 32GB) slot, SSD (Approx. 64GB)
The file for capturing data is limited up to 2GB.
Data saving function Captured data (*11) Built-in RAM, Built-in Flash, SD memory card, SSD (Data is saved directly to it.)
Data in built-in RAM Specified number of data up 2 million samples in increments of 1
Auto save (*11) Available for the built-in RAM
Enabled (ON): Data in the RAM is saved automatically to the built-in Flash, SD memory card, SSD
Disabled (OFF): Data in the RAM is not maintained after power is turned off
Capturing mode (*11) Mode: Off, Normal, Ring, Relay
Ring (*12): Saved most recent data (Number of capturing data: 1000 to 2000000 points, Destination of data: Built-in RAM, Built-in Flash, SD memory card, SSD)
Relay (*13): Saved data to multiple file without losing data until capturing data is stopped (Destination of data: Built-in Flash, SD memory card, SSD)
During data capture (*15) Displaying information in two windows, Hot-swapping the SD memory card, Saving data in between cursors.
Backup (*11) Backup interval: Off, 1, 2, 6, 12, 24 hrs.
Data destination: SD memory card, SSD, FTP server
Data destination for backup cannot be specified to the same storage for destination of capturing data.
Engineering Scale function Measured value can be converted to the engineering unit
Analog voltage: Converts by four reference points (gain, offset)
Temperature: Converts by two reference points (offset)
Pulse count: Converts by two reference points (gain)
Synchronization between units Start and Trigger (*14)
Accuracy of clock (at 23°C) ±0.002% (Monthly deviation approx. 50 sec.)
Operating environment 0 to 45°C, 5 to 85% RH (non condensed)
Power source 100 to 240 V AC, 50 to 60Hz
Power consumption 85VA
Standard accessories Quick guide, CD-ROM, AC power cable
External dimensions (W x D x H) Main module: Approx. 193 x 141 x 160 mm (Excluding Projection),
Alarm output terminal: Approx. 30 x 136 x 145 mm (Excluding projection)
Weight Main module: Approx. 2.2 kg, Alarm output terminal: Approx. 350 g

更多 ELSO Technologies Sdn Bhd 相关资料
ELSO Technologies Sdn Bhd
ELSO Technologies Sdn Bhd Test & Measurement Equipment Distributor Penang, Malaysia, NetAlly, Rigol, Cromwell Distributor Singapore, Selangor, KL, Sarawak ~ ELSO Technologies Sdn Bhd