HI991301 Portable Waterproof pH/EC/TDS Meter (High Range)

HI991301 Portable Waterproof pH/EC/TDS Meter (High Range)

Kategori: Multiparameter Tersedia (Jml:1)
Bisa dibeli di situs resmi
RM 3003.00 Untuk informasi lebih lanjut, kunjungi situs resmi kami di hannamalaysia.com

Deskripsi

The HI991301 is designed to measure pH, conductivity (EC), total dissolved solids (TDS), and temperature with a single probe. From setup to calibration all operations are performed with just two buttons. The meter is waterproof and rated for IP 67 conditions.

The HI12883 pre-amplified multiparameter probe features an easy-to-clean sensor and a cloth junction that can be renewed to increase the life of the probe. To ensure against interference from transient electrical noise, a solid-state amplifier is integrated into the HI12883.

User-selectable features include different TDS factors from 0.45 to 1.00 and a range of temperature coefficients (β) from 0.0 to 2.4% for greater consistency and reproducibility. Standardized buffer recognition values are also selectable.

These instruments easily fit in the palm of your hand and the bottom probe connection ensures the electrode cable doesn’t get in your way. The large, multi-level LCD displays the primary reading, temperature, and calibration guides simultaneously.
 

hi1288 tip

HI12883 Multiparameter Probe

The HI12883 multiparameter probe includes pH, EC/TDS, and temperature measurement. A solid-state pre-amplifier is integrated into the probe to protect against interference from transient electrical noise.

 

  • Amperometric conductivity probe
  • Built-in temperature sensor
  • Easy-to-clean
  • The sensor is easy to clean and maintain by design
  • pH sensor with extendable cloth junction
  • The HI12883 multiparameter probe features a cloth junction that can be pulled longer, effectively increasing the life of the probe

Features at-a-glance

  • 3 sensors in a single probe

pH electrode - Preamplified for resistance to electrical noise including that generated by pumps and ballast

EC/TDS sensor - amperometric two pole design

Temperature - NTC sensor for accurate temperature measurement and compensation

  • Waterproof

Designed to withstand the humidity of a growing environment
 
  • Wide pH Range

pH range is from 0.00 to 14.00 pH
 
  • Selectable EC to TDS conversion factor

Choice of either a 0.5 or 0.7 conversion factor
0.5 (500 CF) conversion factor is for an EC reading of 10 mS/cm = 5 ppt
0.7 (700 CF) conversion factor is for an EC reading of 10 mS/cm = 7 ppt
 
  • Automatic Temperature Compensation
All readings are compensated for variations in temperature
Temperature displayed in oC or oF along with pH, EC or TDS reading
 
  • Stability indicator

Meter displays a clock tag that will disappear when the reading has achieved stability
 
  • HOLD button

Freezes reading on the display to allow recording of measurement
 
  • BEPS (Battery Error Prevention System

Meter will automatically shut off if there is not enough power to get an accurate measurement
 
  • Battery % level at startup
  • Low Battery Indicator
When the level is below 5%, the battery symbol on the LCD blinks to indicate a low battery condition.
 
  • Auto-off
User selectable automatic shut-off ff to conserve battery life. Options are after 8 minutes, 60 minutes , or disabled.
 
SKU HI991301
Product Name Portable Waterproof pH/EC/TDS Meter (High Range) - HI991301
Quote Required Yes
pH Range -2.00 to 16.00 pH; -2.0 to 16.0 pH
pH Resolution 0.01 pH; 0.1 pH
pH Accuracy ±0.02 pH; ±0.1 pH
pH Calibration automatic, at one or two points with two sets of standard buffers (pH 4.01 / 7.01 / 10.01 or pH 4.01 / 6.86 / 9.18)
pH Temperature Compensation automatic 0 to 50 °C; 32 to 122 °F
pH CAL Check (electrode diagnostics) probe condition indicator
pH Sensor Check yes
mV Range ±825 mV (pH‐mV)
mV Resolution 1 mV (pH‐mV)
mV Accuracy ±1 mV (pH‐mV)
EC Range 0.00 to 20.00 mS/cm
EC Resolution 0.01 mS/cm
EC Accuracy ±2% F.S.
EC Calibration automatic, one-point at 12.88 mS/cm
TDS Range 0.00 to 10.00 ppt (g/L)
TDS Resolution 0.01 ppt (g/L)
TDS Accuracy ±2% F.S.
TDS Calibration automatic
EC/TDS Temperature Compensation automatic
Temperature Correction Coefficient β selectable from 0.0 to 2.4 %/ °C with 0.1 increments
EC to TDS Conversion Factor selectable from 0.45 to 1.00 with 0.01 increments
Temperature Range ‐5.0 to 105.0 °C; 23.0 to 221.0 °F
Temperature Resolution 0.1°C; 0.1°F
Temperature Accuracy ±0.5°C; ±1°F
pH/EC Probe HI12883 PVC body, pre-amplified multiparameter probe with internal temperature sensor, quick connect DIN connector and 1m cable
Automatic Shut-Off user selectable
Battery Type/Life 1.5V (3) AAA / approximately 600 hours of continuous use
IP Protection IP67
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight 196 g (6.91 oz.)
Ordering Information HI991301 is supplied complete with HI12883 pH/EC/temperature probe, pH 4.01 calibration buffer sachet (1), pH 7.01 calibration buffer solution sachet (1), HI70030 12.88 mS/cm calibration solution sachet (1), HI70038 6.44 ppt calibration solution sachet (1), HI700601 electrode cleaning solution sachet (2), 100 mL beaker (1), batteries, calibration certificates for both meter and probe, instruction manual, and rugged thermoformed carrying case.



 

Lihat detail lebih lanjut tentang Hanna Instruments (M) Sdn Bhd
Hanna Instruments (M) Sdn Bhd
Hanna Instruments (M) Sdn Bhd Measuring Instruments Selangor, Photometers Supply Penang, Sabah, Test Equipments Supplier Malaysia ~ Hanna Instruments (M) Sdn Bhd