he HI11311 electrode features a built-in microchip that stores sensor type, serial number, and calibration information. It also features a single ceramic junction and the spherical pH sensing portion is made with high temperature glass. This design consideration is ideal for general purpose laboratory use.
The built-in microchip stores sensor type, serial number and calibration information including date, time, offset, slope, probe condition and buffers used. This information is automatically retrieved by edge® once the electrode is plugged in. The ability to transfer information allows for hot swapping of probes without having to recalibrate. All pH measurements are performed within the electrode and transferred digitally to the meter. This overcomes any noise issues associated with the traditional high impedance analog measuring system. Electrical noise can be generated from a built in temperature sensor and while working in a humid environment.
HI11311 features a built-in thermistor temperature sensor that is in the tip of the indicating pH electrode. A thermistor temperature sensor provides for a high accuracy temperature reading and should be as close as possible to the indicating pH electrode in order to compensate for the effect that temperature has on the membrane potential. By having an accurate reading it is possible to provide an accurate temperature compensated reading.
Body Material | glass |
---|---|
Reference | double |
Junction / Flow Rate | ceramic, single / 15-20 μL/h |
Electrolyte | 3.5M KCl |
Range | pH: 0 to 13 |
Max Pressure | 0.1 bar |
Tip Shape | spheric (dia: 9.5 mm) |
Diameter | 12 mm |
Body Length / Overall Length | 120 mm / 195 mm |
Recommended Operating Temperature | -5 to 100°C (23 to 212°F) |
Temperature Sensor | yes |
Matching Pin | yes |
Amplifier | no |
Digital | yes |
Cable | 1 m (3.3') |
Connection | 3.5 mm connector |
Applications | laboratory general purpose |