Conductivity | |||
Model | EC-11 | EC-22 | EC-33 |
Measurement Principle: | 2 Electrode Bipolar AC | ||
Minimum Sample Volume: | 0.12 mL | ||
Measurement Range / Resolution Conductivity: |
Condcutivity 0 to 199 µS/cm (1 µS/cm) 200 to 1999 µS/cm (1 µS/cm) 2.00 to 19.99 mS/cm (0.01 mS/cm) |
Conductivity 0 to 199 µS/cm (1 µS/cm) 200 to 1999 µS/cm (1 µS/cm) 2.00 to 19.99 mS/cm (0.01 mS/cm) 20.0 to 199.9 mS/cm (0.1 mS/cm) |
|
Accuracy: | ±2% full scale (for each range) |
±2% full scale (0 to 19.99 mS/cm) ±5% full scale (20.0 to 199.99 mS/cm) |
|
Maximum Calibration Points: | 2 | 3 | |
Calibration Curves: | 1413 µS/cm, 12.88 mS/cm | 1413 µS/cm , 12.88 mS/cm, 111.8 mS/cm | |
- | TDS Factor (0.4 to 1.0) / EN 27888 / 442 / NaCl | ||
Temperature Display / Resolution: | - | 0 to 50.0°C (0.1°C) | |
Accessories Included: |
1413 µS/cm & 12.88 mS/cm Conductivity Standard Solutions (14 mL each)
|
SALT (NaCl) | |
Model: | Salt-11 |
Measurement Principle: | 2 Electrode Bipolar AC |
Minimum Sample Volume: | 0.12 mL |
Measurement Range / Resolution Conductivity: |
Salt 0.0 to 100.0 g/L (0.1 g/L) Salt 0.0 to 10.0% (0.1%) |
Accuracy: |
±2% full scale (0.0 to 9.9 g/L) ±4% full scale (10.0 to 100.0 g/L) |
Maximum Calibration Points: | 2 |
Calibration Curves: | NaCl / Seawater |
Temperature Display / Resolution: | 0 to 50.0°C (0.1°C) |
Accessories Included: |
0.5% & 5.0% NaCl Standard Solutions (14 mL each)
|
Lihat detail lebih lanjut tentang Environmental Competency Consultancy Sdn Bhd 绿鑫环境顾问有限公司