VIPER measures hyperspectral attenuation and transmission coefficients in the wavelength range of 360 nm and 750 nm, enabling detailed determination of multiple parameters at the same time. The light source is provided by 5 selected, energy-saving LEDs that Warranty a long service life and stable measurement data. VIPER can be used in different media as it is available in multiple path lengths, both in stainless steel or titanium housing.
Light Source
|
5 LEDs
|
|
Detector
|
High-end miniature spectrometer, 256 channels
360 to 750 nm, 2.2 nm/pixel
|
|
Measurement principle
|
Attenuation
|
|
Optical path
|
10 mm, 50 mm, 100 mm, 150 mm, 250 mm
|
|
|
||
Parameter
|
SAC436
Pt-Co color scale (APHA/Hazen) (390 nm, 455 nm)
Colouring based on DIN EN ISO 7887-C (410 nm, 436 nm, 525 nm, 620 nm)
Cr-Co color scale (380 nm, 413 nm)
|
|
Measuring range
|
0.01...2.5 AU (absorption units)
|
|
Measurement accuracy
|
< 0.2 %
|
|
Turbidity compensation
|
Yes
|
|
Data logger
|
~ 2 MB
|
|
T100 response time
|
min. 4 s
|
|
Measurement interval
|
min. 2 s
|
|
|
||
Housing material
|
Stainless steel (1.4571/1.4404) or titanium (3.7035)
|
|
Dimensions (L x Ø)
|
495 mm x 48 mm (with 50 mm path)
|
~ 19.5 x 1.9 (with 50 mm path)
|
Weight stainless steel
|
~ 2.4 kg (with 50 mm path)
|
~ 5.3 lbs (with 50 mm path)
|
Weight titanium
|
~ 1.3 kg (with 50 mm path)
|
~ 2.9 lbs (with 50 mm path)
|
|
||
Interface digital
|
Ethernet (TCP/IP), RS-232 or RS-485 (Modbus RTU)
|
|
Power consumption
|
≤ 1 W
|
|
Power supply
|
12...24 VDC (± 10 %)
|
|
|
||
Maintenance effort
|
≤ 0.5 h/month (typical)
|
|
Calibration/maintenance interval
|
24 months
|
|
System compatibility
|
Modbus RTU
|
|
Warranty
|
1 year (EU: 2 years)
|
US: 2 years
|
|
||
Max. pressure with SubConn
|
30 bar
|
~ 435 psig
|
Max. pressure with fixed
|
3 bar
|
~ 43.5 psig
|
Max. pressure in FlowCell
|
1 bar, 2...4 L/min
|
~ 14.5 psig, 0.5 to 1.0 gpm
|
Protection type
|
IP68
|
NEMA 6P
|
|
||
Probentemperatur
|
+2...+40 °C
|
~ +36 F to +104 F
|
Umgebungstemperatur
|
+2...+40 °C
|
~ +36 F to +104 F
|
Lagertemperatur
|
-20...+80 °C
|
~ -4 F to +176 F
|
Anströmgeschwindigkeit
|
0,1...10 m/s
|
~ 0.33 fps to 33 fps
|
parameter
|
according to
|
unit
|
factor
|
path length [mm]
|
||||
|
|
|
|
10
|
50
|
100
|
150
|
250
|
SAC436
|
DIN EN ISO 7887:2011_method B
|
1/m
|
-
|
1...250
|
0.2...50
|
0.1...25
|
0.06...17
|
0.04...10
|
SAC525
|
DIN EN ISO 7887:2011_method B
|
1/m
|
-
|
1...250
|
0.2...50
|
0.1...25
|
0.06...17
|
0.04...10
|
SAC620
|
DIN EN ISO 7887:2011_method B
|
1/m
|
-
|
1...250
|
0.2...50
|
0.1...25
|
0.06...17
|
0.04...10
|
True Color 410
|
DIN EN ISO 7887:2011_method C
|
mg/L Pt
|
18.52
|
20...3750
|
4...750
|
2...375
|
1.2...250
|
0.8...150
|
Pt-Co-Color 390
|
DIN EN ISO 6271-2016:05
|
mg/L Pt
|
7.4
|
8...1500
|
1.6...300
|
0.8...150
|
0.4...100
|
0.2...60
|
Pt-Co-Color 455
|
DIN EN ISO 6271-2016:05
|
mg/L Pt
|
36.4
|
40...7500
|
8...1500
|
4...750
|
2.4...500
|
1.4...300
|
Cr-Co-Color 380
|
-
|
° (color degree)
|
9.7
|
10.0...2000
|
2...400
|
1...200
|
0.6...130
|
0.4...80
|
Cr-Co-Color 413
|
Gost 3351-74
|
° (color degree)
|
34.1
|
40...7000
|
8...1400
|
4...700
|
2.6...450
|
1.6...275
|