Edge Do

Kategori: Dissolved Oxygen Tersedia
Untuk informasi lebih lanjut, kunjungi situs resmi kami di obsnapinstrument.com

Deskripsi

Hanna Instruments is proud to introduce the world’s most innovative pH meter: edge ®.

edge’s groundbreaking design is the culmination of Hanna’s vision, design capabilities, integrated production and world class R&D. edge is a single meter that can measure pH, Conductivity, and Dissolved Oxygen and is incredibly easy to use. We put a lot of thought into edge’s ease of use.


pH Specifications

pH Range -2.000 to 16.000 pH, -2.00 to 16.00 pH, ±1000 mV
pH Resolution 0.001 pH, 0.01 pH, 0.1 mV
pH Accuracy (@25ºC/77ºF) ±0.2 mV, ±0.01 pH, ±0.002 pH
pH Calibration up to 5 points
pH Calibration Buffers 1.68, 10.01, 12.45, 4.01, 6.86, 7.01, 9.18, two custom buffers
pH Temperature Compensation ATC: -5.0 to 100.0ºC; 23.0 to 212.0°F*

EC Specifications

EC Range 0.00 to 29.99 μS/cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC)**
EC Resolution 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy (@25ºC/77ºF) ±1% of reading (±0.05 μS/cm or 1 digit, whichever is greater)
EC Temperature Compensation ATC (0.0 to 100.0ºC; 32.0 to 212.0 °F), NoTC

TDS Specifications

TDS Range 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy (@25ºC/77ºF) ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
TDS Calibration through EC calibration
TDS Conversion Factor 0.40 to 0.80


Salinity Specifications

Salinity Range 0.0 to 400.0% NaCl, 2.00 to 42.00 PSU, 0.01 to 42.00 PSU, 0.0 to 80.0 g/L*
Salinity Resolution 0.1% NaCl, 0.01 PSU, 0.01 g/L
Salinity Accuracy (@25ºC/77ºF) ±1% of reading

DO Specifications

DO Range 0.00 to 45.00 ppm (mg/L), 0.0 up to 300.0% saturation
DO Resolution 0.01 ppm, 0.1% saturation
DO Accuracy ±1 digit, ±1.5% of reading
DO Calibration one or two points at 0% (HI 7040 solution) and 100% (in air)
DO Temperature Compensation ATC: 0 to 50 ºC; 32.0 to 122.0 °F
Salinity Compenstaion 0 to 40 g/L (with 1 g/L resolution)
DO Altitude Compensation -500 to 4000 m (with 100 m resolution)


Temperature Specifications

Temperature Range -20.0 to 120.0 ºC, -4.0 to 248.0 °F
Temperature Resolution 0.1 °C, 0.1 °F
Temperature Accuracy ±0.5 °C, ±1.0 °F


Additional Specifications

pH Electrode Diagnostics Glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time
Logging 600 records interval logging, stores up to 1000 records: 200 records (log-on-demand and stability logging)
Connectivity 1 micro USB port for charging and PC connectivity, 1 USB port for storage
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Power Supply 5 VDC adapter (included)
Dimensions 202 x 140 x 12.7mm (8” x 5.5” x 0.5”)
Ordering Information HI2040 DO kit includes: HI764080 Dissolved Oxygen electrode, HI7041S Refill electrolyte solution, 2 DO membrane caps, 2 DO membrane cap o-rings and Quality certificate.
Warranty 2 years (probe 6 months)
Notes

* temperature limits will be reduced to actual probe/sensor limits
** with temperature compensation function disabled


Lihat detail lebih lanjut tentang Obsnap Instruments Sdn Bhd
Obsnap Instruments Sdn Bhd
Obsnap Instruments Sdn Bhd Testing Instruments Supplier Malaysia, Measuring Equipments Supply Selangor, Simulation Systems Supplies Kuala Lumpur (KL) ~ Obsnap Instruments Sdn Bhd