The HI96832 is a rugged, portable digital refractometer designed for the measurement of percent volume and freezing point of propylene glycol based solutions. The HI96832 has a high accuracy of ±0.2% volume and ±0.5 °C freezing point and features simple operation that gives reliable results each and every time. All readings are automatically compensated for temperature variations and displayed with a 1.5 second response time. The sealed flint glass prism and stainless steel well are easy to clean. Just wipe with a soft cloth in preparation for the next sample.
The HI96832 Digital Refractometer utilizes the measurement of refractive index to determine the percent volume and freezing points of propylene glycol based solutions. Propylene glycol is used in glycol chillers and dispensing systems intended for food applications, such as beer and wine. Maintaining the appropriate percent of glycol is important in glycol chillers as too low of a percentage could lead to the system freezing and rupturing while too much glycol will reduce the systems efficiency. The instrument converts the refractive index of a sample to percent volume or freezing point. These conversions are based on internationally recognized references for unit conversion and temperature compensation for glycol solutions (e.g. CRC Handbook of Chemistry and Physics, 87th Edition). The HI96832 is an easy-to-use tool for measuring freezing point and percent volume in propylene glycol solutions, on the production floor or in the lab.
Freezing point is displayed as a temperature from 0.0 to -51.0 °C corresponding to 0-60 % by volume. When the display shows “LO”, the freezing point is at a minimum (below -51 °C).
To make an X % Propylene Glycol Solution, follow the procedure below:
Propylene Glycol | Total Volume | Expected Freezing Point Value | |
10% V/V | 10.00 mL | 100.00 mL | -3.4 °C (25.9 °F) |
40% V/V | 40.00 mL | 100.00 mL | -21.3 °C (-6.3 °F) |
The HI96832 takes measurements based on a sample’s refractive index. Refractive index is a measurement of how light behaves as it passes through the sample. Depending on the sample’s composition, light will refract and reflect differently. By measuring this activity with a linear image sensor, the sample’s refractive index can be assessed and used to determine its physical properties such as concentration and density. In addition to the linear image sensor, the HI96832 uses an LED light, prism and lens to make measurement possible.
Variations in temperature will affect the accuracy of refractometry readings, so the use of temperature compensation is highly recommended for reliable results. The HI96832 contains a built-in temperature sensor and is programmed with temperature compensation algorithms in accordance with internationally recognized references which vary based on the parameter being measured.
SKU | HI96832 |
---|---|
Product Name | Digital Refractometer for Propylene Glycol Analysis - HI96832 |
Quote Required | Yes |
Propylene Glycol Range | 0 to 100% |
Propylene Glycol Resolution | 0.1 % |
Propylene Glycol Accuracy | ±0.3 % |
Temperature Range | 0 to 80°C (32 to 176°F) |
Temperature Resolution | 0.1°C (0.1°F) |
Temperature Accuracy | ±0.3 °C (±0.5 °F) |
Temperature Compensation | automatic between 0 and 40°C (32 to 104°F) |
Automatic Shut-Off | after three minutes of inactivity |
Battery Type/Life | 9V / approximately 5000 readings |
Dimensions | 192 x 102 x 67 mm (7.6 x 4.01 x 2.6”) |
Weight | 420 g (14.8 oz.) |
Refractometer Light Source | yellow LED |
Minimum Sample Volume | 100 μL (to cover prism totally) |
Sample Cell | stainless steel ring and flint glass prism |
Measurement Time | approximately 1.5 seconds |
Ordering Information | HI96832 is supplied with battery and instruction manual |
Notes | Enclosure Rating |