When equipped with various equipment packages, the tube furnace product lines can be adapted for operation with nonflammable or flammable gasses or for vacuum operation.

Gas Supply System 1 for non-flammable protective or reactive gases Not gastight, no vacuum operation

This package represents a basic version sufficient for many applications, for operation with non-flammable protective or reactive gasses. The standard working tube made of ceramic C 530 delivered with the furnace can still be used.

Gas Supply System 2 for non-flammable protective or reactive gases/vacuum operation

For increased atmospheric purity requirements in the working tube, we recommend this gas supply system. The system can also be equipped for vacuum operation.

Vacuum Operation

Gas Supply System 4 for hydrogen, fully-automatic, unattended operation

Adding gas supply system 4 to the tube furnace allows operation under a hydrogen atmosphere. During hydrogen operation, a safety pressure of approx. 30 mbar is ensured in the working tube. Surplus hydrogen is burnt off in an exhaust gas torch. With extended safety logic and an integrated nitrogen purge container, the system can be used for fully-automatic, unattended operation. Equipped with a Safety-PLC control system, pre-purging, hydrogen inlet, operation, fault monitoring and purging at the end of the process are carried out automatically. In case of default, the tube is immediately purged with nitrogen and the system is automatically switched to a safe status.

  • Available for tube furnaces RSH, RSV, RSRC, RHTH and RHTV
  • Safety system for operation with flammable gases including monitoring of torch function and overpressure
  • Extended safety control system with emergency tube purging in case of default
  • Emergency purge container
  • Safety-PLC control system with touchpanel for data input
  • Exhaust gas torch
  • Pressure switch for monitoring the safety pressure
  • Gas supply system for H2 and N2. Volume adjustment is carried out by hand (the customer provides an H2 supply at 1 bar, an N2 supply at 10 bar, an O2 supply at 6-8 bar and a propan supply at 300 mbar)