The insulation, load and temperature monitor or insulation monitoring device (IMD) ensures absolute reliability of measurement by means of a codified signal that is able to detect the disturbances generated by the common devices in the operating theatres/rooms and critical care areas; this avoids unwanted alarm signaling.

The HRI-R40 model measures insulation resistance, insulation impedance, load and temperature of the medical isolation transformer.
 

Technical Features:

Power Supply AC 230V ± 20%, 50...60 Hz
Monitoring Network AC 24...230V, 50...60 Hz
Maximum Consumption 6 VA
Max. Measuring Voltage / Current <DC 24 V / <1 mA
Control Signal Composed Codified Signal
Insulation Voltage 2.5kV / 60 sec
Signalations Low insulation resistance 50...500 kOhm
Low insulation impedance 50...500 kOhm
Overtemperature T1/T2 0...200°C
Overload 1...99.9 A
Error / link fail
Outputs Auxiliary relay NO-C-NC 5A AC 250V
Voltage Supply for Signalling DC 24V
RS 485 Modbus RTU
Memory MAX/MIN Parameters
Protection Degree Frontal-IP 50, enclose-IP 20
Dimensions (W x D x H) / Weight 105mm x 74mm x 90mm / 500g
Standards / Certifications IEC 60364-7-7-10, IEC/EN 61557-8, EN 61326-1, CE, ABE