Description
DX 310/311 is a digital combined temperature and relative humidity probe for the PRO D01 / PRO D05 handheld meters.
The probe supplies the following calculated quantities:
-
Dew point
-
Wet bulb temperature
-
Absolute and specific humidity
-
Mixing ratio
-
Specific enthalpy
-
Partial vapor pressure
-
Frost point temperature
-
Saturation vapor pressure above water and above ice
The probe is supplied factory-calibrated in multiple points for relative humidity and it is ready to use.
The probe can be optionally supplied with a DAkkS/ACCREDIA calibration certificate.
The calibration data are stored in the probe, allowing for interchangeability without the need for recalibration.
Equipped with 4-pin M12 connector and 2, 5 or 10 m cable (only DX311).
-
Output:
-
UART (TTL 3.3V): DX-Sensor-Interface
-
Power supply:
-
3.3…6 Vdc
-
Power consumption:
-
<1 mA typ.
-
Connection:
-
4-pole M12
-
Dimensions:
-
Stem: Ø14 x 101 mm Overall: Ø14 x 114.8 mm
-
Weight:
-
20 g approx.
-
Material:
-
Stem and protector cap: PBT
-
Sensor:
-
RH: Capacitive
Temperature: Pt100
-
Measuring range:
-
RH: 0 ... 100 %
Temperature: -50 ... +160 °C
-
Resolution:
-
RH: 0.01 %
Temperature: 0.01 °C
-
Accuracy:
-
RH: ± 1.2% (0…85%) / ±2% (85…100%) @ T=0…50 °C (1.5 + 1.5% of the measured value)% @ T=remaining range
Temperature: ± 0.1 °C ± 0.1% of the measured value
-
Operating conditions:
-
-40…+80 °C / 0…100 %RH
-
Long-term drift:
-
RH: ± 0.5 %RH/year
Temperature: ± 0.03 °C/year
-
RH response time:
-
10 s (10 → 80 %RH; air speed=2 m/s @ constant temperature)
-
Calculated quantities:
-
Dew point Wet bulb temperature Absolute humidity Specific humidity Mixing ratio Specific enthalpy Partial vapor pressure Frost point temperature Saturation vapor pressure above water Saturation vapor pressure above ice