Measurement line type | Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire, three-phase 4-wire |
---|---|
Number of input channels |
Max. 6 channels; each input unit provides 1 channel for simultaneous voltage and current input (Voltage measurement unit: Photoisolated input, resistance voltage divider, Current measurement unit: Isolated input from current sensor) |
Measurement items | Voltage (U), current (I), active power (P), apparent power (S), reactive power (Q), power factor (λ), phase angle (φ), frequency (f), efficiency (η), loss (Loss), voltage ripple factor (Urf), current ripple factor (Irf), current integration (Ih), power integration (WP), voltage peak (Upk), current peak (Ipk) |
Harmonic measurement: Harmonic active power, select calculation order from 2nd order to 100th order |
|
Waveform recording: Voltage and current waveforms/ Motor pulse: Always 5 MS/s Motor waveforms: Always 50 kS/s, 16 bits Recording capacity: 1 Mword × ((voltage + current) × number of channels + motor waveforms) |
|
Motor analysis (PW6001-11 to -16 only): Voltage, Torque, Rotation, Frequency, Slip, or Motor output | |
Measurement range |
Voltage range: 6 to 1500 V, 8 ranges Current range (Probe 1) : 400 mA to 1 kA (depends on current sensor) Current range (Probe 2) : 100 mA to 50 kA (depends on current sensor) Power range: 2.40000W to 4.50000MW (depends on combination of voltage and current range) Frequency range: 0.1 Hz to 2 MHz |
Basic accuracy |
Voltage: ±0.02 % rdg. ±0.02 % f.s. Current: ±0.02 % rdg. ±0.02 % f.s. + current sensor accuracy Active power: ±0.02 % rdg. ±0.03 % f.s. + current sensor accuracy |
Synchronization frequency range |
Power measurement: 0.1 Hz to 2 MHz Harmonic measurement: 45 Hz to 66 Hz (IEC standard mode), 0.1 Hz to 300 kHz (Wideband mode) |
Frequency band | DC, 0.1 Hz to 2 MHz |
Data update rate |
Power measurement: 10 ms/ 50 ms/ 200 ms Harmonic measurement: 200 ms (IEC standard mode), 50 ms (Wideband mode |
Data save interval |
OFF, 10 msec to 500 msec, 1 sec to 30 sec, 1 minute to 60 minutes User-selected from all measured values, including harmonic measured values, Specified measured values can be saved in internal memory or USB flash drive |
External interfaces | USB (memory), LAN, GP-IB, RS-232C (for communication/LR8410 link), External control ,Synchronization control |
Logger connectivity | Sends measured values wirelessly to logger by using a Bluetooth® wireless technology serial conversion adapter. (Supported devices: Hioki LR8410 Link-compatible loggers), Ver. 2.0 and later |
Power supply | 100 to 240 V AC, 50/60 Hz, 200 VA max |
Dimensions and mass | 430 mm (16.93 in)W × 177 mm (6.97 in)H × 450 mm (17.72 in)D, 14 kg (49.4 oz) (PW6001-16) |
Accessories | Instruction Manual ×1, Power cord ×1, D-sub connector × 1 (PW6001-1x only) |