Color spectrophotometer CS-520 greatly expands the application boundary of color spectrophotometer; automatic calibration, magnetic seat charging, touch screen operation, switchable apertures, UV included / excluded... it will fully upgrade the user exper.
Technical Data
Model | Spectrophotometer CS-520 |
Geometry |
d/8(diffused illumination,8° viewing) SCI(Specular Component Included),SCE(Specular Component Excluded) |
Light Source | LED(Full band balanced LED) |
Color Spaces and Indices | Reflectance, CIE-Lab, CIE-LCh, HunterLab, CIE Luv, XYZ, Yxy, RGB, Color difference(ΔE*ab, ΔE*cmc, ΔE*94,ΔE*00),WI(ASTM E313-00,ASTM E313-73,CIE/ISO, AATCC, Hunter, Taube Berger Stensby), YI(ASTM D1925,ASTM E313-00,ASTM E313-73), Blackness(My,dM),Color Fastness, Tint,(ASTM E313-00),Color Density CMYK(A,T,E,M), Milm, Munsell, Opacity, Color strength |
Illuminants | A,B,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,U35,DLF,NBF,TL83,TL84 |
Observer Angles | 2°,10° |
Sensor | 256 pixel CMOS sensor |
Calibration | Auto |
Wavelength Interval | 10nm |
Wavelength Range | 400-700nm |
Reflectance Range | 0-200% |
Reflectance Resolution | 0.01% |
Sphere Size | 40mm |
Repeatability |
Chromaticity value: Max. :dE*ab< 0.04 Average :dE*ab< 0.03 Standard deviation within ΔE*ab 0.02 (when a white tile is measured 30 times at 5-seconds interval) Reflectance: Standard deviation < 0.1% |
Illumination Area/Aperture |
LAV:Φ8mm/Φ11mm MAV:Φ4mm/Φ6mm MINI:1*3mm |
Inter-instrument agreement |
ΔE*ab<0.2 (BCRA Series Ⅱ, average measurement of 12 tiles,MAV/SCI) |
Test Time | About 1s |
Storage | APP mass storage |
Battery | Rechargeable, 8000 times continuous tests, 3.7V/3000mAh |
Screen | IPS Full Color Screen,2.4 inches |
Interface | USB,Bluetooth |
Lamp Lifetime | 10 years, 1 million tests |
Language | Chinese and English |
Software | Android, IOS, Windows |
Pass Metrology | Yes, metrology level I |
Camera | with |
Instrument Size | L x W x H (77.8mm x 53.2mm x 185.7mm) |
Weight | About 300g |