F701-S is a weighing indicator with high specifications and various functions such as super-low-drift high-precision amplifier, fast-sampling, weighing sequence functions that are useful for automatic weighing machines and control I/O.
Analog|Setting|External signal|Interface|Display|General specification|Attachment
Analog | |
Excitation voltage | DC5V±5% Output current:within 90mA Ratio metric type (Up to six 350Ω load cells can be connected in parallel.) |
---|---|
Signal input range | -0.5~3.0mV/V , 0 to 3.0mV/V (when OIML R76-1-compliant) |
Zero adjustment range |
Automatic adjustment by digital operation -0.2~2.0mV/V , 0 to 2.0mV/V (when OIML R76-1-compliant) |
Span adjustment range |
Automatic adjustment by digital operation |
Minimum input sensitivity | 0.15μV/count , 0.5μV/count (when OIML R76-1-compliant) |
Accuracy |
|
A/D converter |
Conversion rate…300 times/sec. Conversion resolution…24bit(binary) |
Minimum indication resolution | 1/10000, 1/6000 (when OIML R76-1-compliant) |
Secondary calibration | Equivalent input calibration Minimum indication resolution for secondary calibration:1/1000(ordinary temperature) |
Setting | |
Setting method |
Settings are made by operating the membrane keys. Also, settings can be made from a host computer through the RS-485 interface. |
---|---|
Memory of set value |
Calibration value and a part of set value: NOV.RAM (nonvolatile RAM) Other set values: F-RAM (nonvolatile RAM) |
Protect of set value | Protect can be set by Lock switch and Lock parameter. |
Setting item |
Upper limit/Lower limit/Near zero/Set point 1/Set point 2/Compensation/Over/Under/Final/Comparison inhibit time/Judging time/Complete output time/Compensation feeding time/Auto free fall compensation regulation/Weighing function 1/Weighing function 2/Tare setting/Tare display/Digital low pass filter/Moving average filter/MD/ZT period/ZT range/Total comparison selection/Total limit (most significant 4 digits)/Total limit (least significant 5 digits)/Count limit/Weighing function 3/Key invalid・Lock/Input selection 1/Input selection 2/Output selection 1/Output selection 2/Error output selection/Reserve output selection/Password/Bag clamp output time/Discharging time/Weighing start time/AZ times/Judging times/Sequence mode 1/Sequence mode 2/Filling enhancement value/Expanded function selection 1/Expanded function selection 2/Balance weight value/Capacity/Min. scale division/DZ regulation value/Display selection 1/Gravitational acceleration/Net over/Gross over/Display selection 2/RS-485 I/F setting/RS-485 Communication type/RS-485 ID/Transmission delay time/Zero calibration/Span calibration/Equivalent calibration/Option display |
External signal (You can specify whether PNP (Source) type or NPN (Sink) type when order the F701-S.) | |
Output signals(16points) |
SP1/SP2/SP3/Complete/Discharge/Bag clamp /Error selection 1,2/Output selection 1~6/Reserve 1,2 At signal ON, output transistor ON. ※External voltage must be prepared separately by customer. |
---|---|
Input signals (16points) |
Operation permission/Weighing start/Stop/Discharge command/Manual discharge/Discharge gate open/Accumulation clear/Input selection 1~6/Code 1,2,4 Contact (relay, switch etc.) or non-contact (transistor, open collector etc.) can be connected. ※External voltage must be prepared separately by customer. |
Interface | |
RS-485 communication interface (Select from Modbus-RTU and original format) 232:RS-232C communication interface(option) BCO:BCD parallel data output interface (sink type) (option) BSC:BCD parallel data output interface (source type) (option) DAC:D/A converter (option) ODN:DeviceNet interface (option) PRF:PROFIBUS interface (option) ※PROFIBUS-DPV0 CCL:CC-Link interface (option) |
Display | |
Display |
18.5mm in character height, Numerical display on LCD(7 digit) Sub display:7.3mm in character height (14 digit) |
---|---|
Indicated value | 5 digit sign: negative display at the highest digit |
Accumulation value | 9 digit ※This can be changed to ''Accumulation count (4 digit)'', ''Final(5 digit)'', ''Code(1 digit)'', ''Total discharge count (6 digit)'' and ''Discharge count (5 digit)''. |
Display frequency | Select from 1, 2, 5, 10, 20 times/sec.(The system speed is 300 times/sec.) |
Capacity | 5 digit |
Min. scale division |
Setting allowable in the range of 1~50. * Only the following settings are valid when OIML R76-1-compliant: 0.001, 0.002, 0.005, 0.010, 0.020, 0.050, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50 |
Decimal point | Select from 0, 0.0, 0.00, 0.000 |
Over display |
LOAD:A/D converter input over, OFL1:Net Over, OFL2:Capacity +9 scale division, -OFL2:Capacity -over or -20 scale division over, OFL3:Gross Over |
Center zero |
A true zero point or the center of each value is displayed. *Only true zero point is displayed when OIML R76-1-compliant. |
Status display |
CLAMP/SP3/SP2/SP1/HOLD/ COMPL./ZT/ZALM/STAB/TARE/NET/GROSS/NZ/ D.CHG/HI LIM/HI/GO/LO/LO LIM/CZ/LOCK |
General specification | |
Power supply voltage | AC100~240V(+10%-15%) (free power source 50/60Hz) |
---|---|
Power consumption | 15W max. |
Inrush current |
2A, 3mSec:AC240V average load condition (cold start at room temperature) |
Operating conditions |
Temperature…Operation -10℃~+40℃ Storage -20℃~+85℃ Humidity…85% or less (non-condensing) |
External dimensions | 192(W)×96(H)×140(D) mm (Projections excluded) |
Panel cutout dimensions |
186 +2 -0(W)×92 +1 -0(H)mm ※Installation panel thickness: 1.6~3.2mm |
Weight | Approx. 2kg |
Attachment | |
AC input cord (Nominal rating 125V) 2m | 1 |
---|---|
Loadcell connector | 1 |
Load cell connector rubber | 1 |
External I/O signal connector | 1 |
Operation manual | 1 |
Rubber packing | 1 |
BCD output connector(with BCD output option) | 1 |
DeviceNet connector(with DeviceNet option) | 1 |
CC-Link connector(with CC-Link option) | 1 |
Model | Interface |
232 | RS-232C communication interface |
---|---|
BCO | BCD parallel data output interface (sink type) |
BSC | BCD parallel data output interface (source type) |
DAC | D/A converter |
ODN | DeviceNet interface |
PRF | PROFIBUS interface |
CCL | CC-Link interface |
Product | Description |
CAAC3P-P2 | AC input cord 2m |
---|---|
CAAC3P-CEE7/7-P1.5 | AC input cord(voltage resistance:250V)1.5m |
CA4131 | (6-wired)cable with JRC connector at one end 3m |
CA4230 | JRC-PRC (6-wired)conversion relay cable 0.3m |
CA4311 | JRC-PRC (6-wired)conversion relay cable (4-wired to 6-wired)(for 520A use)1m |
CN3P-2P | 3P-2P converter plug for AC input cord |
CN10 | Loadcell connector |
CN34 | D-Sub 9p connector for RS-232 |
CN51 | BCD output connector |
CN52 | FCN series I/O connector (with cover) |
CN57 | FCN series I/O connector (with diagonal cover) |
CN71 | CC-Link connector |
CND01 | DeviceNet connector |