General Specification Physical |
|
---|---|
Dimensions (hxwxd) | 470 x Ø80 mm |
Weight | 2.2 kg (inc 10 metres of cable) |
Protection Class | IP68 |
Enclosure Material | Black Acetal Co-polymer |
Cable Entries | Integral Cable Gland |
Wetted Parts | Black Acetal, 316 Stainless Steel, Glass, Polyurethane, Nitrile |
Cable Type | 4 core, 2 Twisted Pair, 9 mm O/D Polyurethane Coated |
Cable Length | 10 metres standard, 100 metres maximum |
Service Requirement |
Automatic Self Cleaning Seal Service every 3500 cleans (application dependent) |
Environmental Data | |
Operating Temperature | 0 to 50ºC |
Interface to Monitor | |
Power Supply | 12 VDC from Monitor |
Sensor Output | ModTechw², RS485 |
Measurement Characteristics | |
---|---|
Range | 0 - 20,000 mg/l |
Accuracy | ±5% of reading or 100 mg/l whichever is greater |
Limit of Detection | 100 mg/l |
Minimum Operating Range | 0 - 2,000 mg/l |
Measurement Principle | Light Attenuation |
Wavelength | 860 nm Infrared |
Pressure Rating (Depth) | 1Bar or 14.5 psi |
Flowrate | Not affected by flowrate, avoid deadspots and extreme turbulence |
Units | User Selectable mg/l, g/l, ppm, %SS, NTU |
The actual measuring range will depend on the nature of the sample being monitored, for processes normally operating at 10,000 mg/l or more should consider use of the HR version of this sensor. Processes that use carbon will provide higher than expected readings and the HR version of the sensor should be used. |