Input : Capacitive humidity sensor & Pt100
Output : 4 ... 20 mA / 0 ... 10 V / RS-485
Range : -40°C ... +180°C
Customized up to 200°C (Optional)
0 ... 100%RH (non-condensing)
Accuracy : [Temperature]±0.15°C+0.002°C x tactual
[Humidity (0 ... 90%)]Nonlinear error:±1.2%RH, Hysteresis error:±0.8%RH, Repeatability error:±0.4%RH
[Humidity (90 ... 100%)]±2%RH
IP rating : IP67
●YUDEN-TECH CO.,LTD. Calibration Laboratory - ( ILAC / TAF ) Test report, for more detailed information please contact us.
(TAF accreditation:3032, complying with ISO / IEC 17025 ) TAF has mutual recognition arrangement with ILAC MRA
Description
Adaptation to high Temp. and high humid. conditions or harsh environments such as chemical pollution & condensation
IP67, Stainless steel SUS304, Pressure resistance 16 bar
330° rotatable, lightweight, unrestricted installation
Optional on-site real-time display
Process temperature up to 180°C, custom up to 200°C (Optional)
Output:Analog / RS-485
Wide measurement range, Temp. : -40 ... +180°C, Humidity : 0 ... 100%RH
Measuring parameter:relative humidity, dew point temperature, frost point temperature, wet bulb temperature, water vapor partial pressure, mixing ratio, absolute humidity and enthalpy.
THM06 has a sturdy outer housing and a wide range of applications. It is also suitable for use in harsh environmental conditions.
Capacitive sensor, high accuracy of temperature, and humidity measurement, fast response, fast recovery after condensation.
It is stability even in long-term high humidity, also it has temperature compensation and computer linear calibration temperature and humidity function, suitable for industrial process gas supply or consumption, compressed dry air flow, and consumption monitoring.
Specification
Input
Signal type : Capacitive humidity sensor & Pt100
Temperature range : -40°C ... +180°C, Customized up to 200°C (Optional)
Humidity range : 0 ... 100%RH (non-condensing)
(Please refer tophysical quantity measuring range list)
Output
Signal : 4 ... 20 mA / 0 ... 10 V / RS-485
Signal connection : 3-wire
Accuracy
Temperature : ±0.15°C+0.002°C x tactual
Humidity (0 ... 90%) : Nonlinear error:±1.2%RH, Hysteresis error:±0.8%RH, Repeatability error:±0.4%RH
Humidity (90 ... 100%) : ±2%RH
Reaction time t90 (at +25°C) : <20 sec (S.S. metal grid filter with mesh), <30 sec (Sintered filter)
Load resistance : Current output:≦ 500 Ω, Voltage output:≧ 10 KΩ
Electrical
Power supply : DC 24 V±10%
Current consumption : DC 24 V:60 mA
Electrical connections : M12 metal connector
Environmental
Medium : Air, or medium compatible with stainless steel
Temp. Operating : -20 ... +60°C (Body);-40 ... +180°C (Probe)
Humid. Operating : 0 ... 90%RH (Body, non-condensing)
Probe pressure : 16 bar
Storage temperature : -20 ... +60°C
Installation
Installation : Metal connector
Protection
IP rating : IP67
Electrical protection : ■Polarity reversal protection ■Over-voltage ■Short-circuit
Material
Housing : SUS304
Weight : Metal:318 g (without wire)
Application