Thermocouple, Resistance Thermocouple
3126 | - | □ | - | □ | - | □ |
① | ② | ③ |
Code | Type of T/C | Measuring Range | Display Range | Accuracy *1 |
---|---|---|---|---|
01 | R | 100 to 1700°C | -50 to 1750°C | ±(0.5% of rdg + 1°C) |
02 | K | -100 to 1300°C | -100 to 1350°C | |
04 | J | 0 to 1200°C | -50 to 1250°C | |
05 | T | -100 to 400°C | -150 to 420°C |
Code | Type of sensor | Measuring Range | Display Range | Accuracy *2 |
---|---|---|---|---|
12 | Pt100Ω | -199.9 to 199.9°C | -199.9 to 199.9°C | ±(0.2% of rdg + 0.3°C) |
11 | -199 to 600°C | -199 to 650°C | ±(0.4% of rdg + 1°C) |
Display | Red or green LED (character height 10 mm) with zero-suppress function. |
---|---|
Over-range indication | When exceeding the display range, flashes with minimum or maximum value. |
Open circuit of input |
Thermocouple: flashing with minimum value Resistance thermometer: flashing with maximum value |
Resolution |
Thermocouple: 1°C Resistance thermometer: 1°C or 0.1°C |
External resistance |
Thermocouple: Less than 500 Ω Resistance thermometer: Less than 5 Ω/lead wire |
Overload | DC ±3.3 V |
Sampling rate | 2.5 times/s |
Input configuration | Single ended, floating input |
A/D conversion | ⊿-Σ conversion system |
Noise rejection | Normal mode (NMR) More than 50dB |
Hold function | Holds measured data and analog output (option). (Not isolated from the input) |
°F indication | °F = °C × 9 / 5 + 32.0 |
Withstand voltage |
Between input terminals and external case: AC 500 V each for 1min. Between power supply terminals and external case: AC 500 V each for 1min. Between power supply terminals and input terminals: AC 500 V each for 1min. |
Insulation resistance | DC 500 V More than 100 MΩ |
Power supply | DC 12 to 24 V |
Tolerance of source voltage | DC 9 to 32 V |
Power consumption |
DC 12 V … 60 mA DC 24 V … 40 mA |
Operating temperature | 0 to 50°C |
Storage temperature | -20 to 70°C |
Weight | Approx. 55 g |
Mounting Method | Snap-in type |