Thermocouple, Resistance Themometer
3155C | - | □ | - | □ |
① | ② |
Code | Type | Measuring Range | Display Range | Accuracy |
---|---|---|---|---|
TC | R | 100 to 1700°C | -50 to 1750°C | ±(0.3% of rdg + 1°C) |
K | -100 to 1300°C | -199 to 1350°C | ||
E | -130 to 1000°C | -199 to 1050°C | ||
J | -140 to 1200°C | -199 to 1250°C | ||
T | -199 to 400°C | -199 to 420°C | ||
B | 600 to 1800°C | -20 to 1810°C | ||
N | -100 to 1300°C | -199 to 1350°C |
Code | Type | Measuring Range | Display Range | Accuracy |
---|---|---|---|---|
PT | Pt100Ω | -199.9 to 600.0°C | -199.9 to 650.0°C | ±(0.2% of rdg + 0.3°C) |
JP100Ω |
Display | Red LED, character height 51 mm, with zero-suppress function. |
---|---|
Over-range display | When exceeding the display range, flashes with minimum or maximum value. |
Open circuit of input |
Thermocouple: flashs with min. value of display range Resistance Thermometer: flashs with max. value of display range |
Resolution |
Thermocouple: 1°C Resistance Thermometer: 0.1°C |
External resistance |
Thermocouple: Less than 500 Ω Resistance thermometer: Less than 5 Ω/lead wire |
Overload | DC ±10 V |
Sampling rate | Approx. 2.5 times/sec. |
Input configuration | Single ended, floating input |
A/D conversion | ⊿-∑ conversion system |
Noise rejection |
Normal mode (NMR) More than 50 dB Common mode (CMR) More than 110 dB Power source line mixing noise 1000 V |
Hold function | Holds measurement data. (Not isolated from input) |
Withstand voltage |
Between input terminal and external case: AC 1500 V for 1min. Between power supply terminals and external case: AC 1500 V for 1min. Between power supply terminals and input terminals: AC 1500 V for 1min. |
Insulation resistance | DC 500 V More than 100 MΩ |
Operating temperature | 0 to 50°C |
Weight | Approx. 1.3 kg |