WINTERS LE3 GENERAL PURPOSE PRESSURE TRANSMITTER

WINTERS LE3 GENERAL PURPOSE PRESSURE TRANSMITTER

Kategori: PRESSURE TRANSMITTERS Tersedia
Untuk maklumat lanjut, lawati laman web rasmi kami di improtechonline.com

Penerangan

APPLICATIONS:

DESCRIPTION & FEATURES:


SPECIFICATIONS
Output Signal

Standard 2-wire: 4-20 mA / Vs = 8-32 Vdc

Optional 3-wire: 0-10 V / Vs = 14-30 Vdc

Optional 3-wire ratiometric: Vsig = 0.5-4.5 Vs / Vs = 5 ± 0.5 Vdc

Accuracy ±0.5% FSO
Permissible Load

2-wire: Rmax = [(Vs – Vs min) / 0.02] Ohm

3-wire: Rmin = 10 kOhm

Influence Effects

Supply: 0.05% FSO / 10 V

Load: 0.05% FSO / kOhm

Response Time

2-wire: <10 ms

3-wire: <3 ms

Measuring Rate 1 kHz
Thermal Error <±0.5% FSO / 10K
Permissible Temperatures

Medium: -13°F to 257°F (-25°C to 125°C)

Electronics environment: -13°F to 185°F (-25°C to 85°C)

Storage: -40°F to 185°F (-40°C to 85°C)

Short-circuit Protection

Permanent

3-wire ratiometric: None

Reverse Polarity Protection No damage. No function.
Electromagnetic Protection Emission and immunity according to EN 61326
Stability

Vibration: 10 g, 25 Hz to 2 kHz according to DIN EN 60068-2-6

Shock: 500 g / 1 ms according to DIN EN 60068-2-27

Pollution Degree 4, electrical equipment for outdoor use
Pressure Port / Housing 304 SS
Wetted Parts

Seals: FKM (standard)

Diaphragm: Ceramic AI2O3 96%

Weight 120 g
Current Consumption

2-wire: 25 mA max

3-wire voltage: 5 mA (short-circuit current 20 mA)

3-wire ratiometric: 1.5 mA

Long-term Stability <±0.3% FSO/yr at reference conditions
Operation Life >100 × 106 cycles
CE Conformity

EMC directive: 2004/108/EC

Pressure equipment directive: 97/23/EC (module A). Valid for devices with max permissible overpressure > 3,000 psi

 
 

Lebih maklumat tentang TECH IMPRO AUTOMATION SOLUTION SDN BHD
TECH IMPRO AUTOMATION SOLUTION SDN BHD
TECH IMPRO AUTOMATION SOLUTION SDN BHD Industrial Automation Supplier Malaysia, Flow Meter Supply Penang, Pressure Transmitter Supplies Butterworth ~ TECH IMPRO AUTOMATION SOLUTION SDN BHD