Fluke Insulation Testers - Fluke 1587 and 1577 Insulation Multimeters

Fluke Insulation Testers - Fluke 1587 and 1577 Insulation Multimeters

Kategori: FLUKE Stok Habis
Untuk maklumat lanjut, lawati laman web rasmi kami di pacificradio.com.my

Penerangan

Specifications
Voltage DC
Maximum Voltage:  1000 V
Accuracy:  1577: ±(0.2%+2)
   1587: ±(0.09%+2)
Maximum Resolution:  0.001V
Voltage AC
Maximum Voltage:  1000 V
Accuracy:  1577: ±(2%+3)
   1587: ±(2%+3)
AC Bandwidth:  1587: 5kHz*
   *with low pass filter; 3db @ 800 Hz
Maximum Resolution:  0.1 mV
Current DC
Maximum Amps:  400mA
Amps Accuracy:  1577: ±(1.0%+2)
   1587: ±(0.2%+2)
Maximum Resolution:  0.01 mA
Current AC
Maximum Amps:  400mA
Amps Accuracy:  1577: ±(2.0%+2)*
   1587: ±(1.5%+2)*
   * 1kHz bandwidth
Maximum Resolution:  0.01 mA
Resistance
Maximum Resistance:  50 MΩ
Accuracy:  1577: ±(1.2%+2)
   1587: ±(0.9%+2)
Maximum Resolution:  0.1Ω
Capacitance(1587 only)
Maximum Capacitance:  9,999 µF
Accuracy:  ±(1.2%+2)
Maximum Resolution:  1 nF
Frequency (1587 only)
Maximum Frequency:  100 kHz
Accuracy:  ±(0.1%+1)
Maximum Resolution:  0.01 Hz
Temperature Measurement (1587 only)
   -40.0 °C to 537 °C
   -40.0 °F to 998 °F
   excluding probe
Diode Test (1587 only)
Range:  6 V
Resolution:  1 mV
Accuracy:  ±(2 % + 1)

Insulation Test
Minimum Test Current @ 1kΩ/V:
   1mA
Test Voltages:
1577:  500, 1000 V
1587:  50, 100, 250, 500, 1000 V
Maximum Resistance per Test Voltage:
50V:  
100V:  
250V:  
500V:  
1000V:  1577: 600 MΩ
   1587: 2 GΩ
Maximum Resolution per Test Voltage:
50V:  0.01 MΩ
100V:  0.01 MΩ
250V:  0.1 MΩ
500V:  0.1 MΩ
1000V:  0.1 MΩ
Accuracy per Test Voltage:
50V:  ±(3 % + 5)
100V:  ±(3 % + 5)
250V:  ±(1.5 % + 5)
500V:  1577: ±(2.0 % + 5)
   1587: ±(1.5 % + 5)
1000V:  1577: ±(2.0 % + 5)
   1587: ±(1.5 % + 5) to 600 MΩ, ±(10 % + 3) above 600 MΩ

Environmental Specifications
Operating temperature
   –20 °C to +55 °C
Storage temperature
   -40°C to +60°C
Humidity (Without Condensation)
   0% - 95% (10 °C – 30 °C)
   0% - 75% (30 °C – 40 °C)
   0% - 40% (40 °C – 55 °C)
Operating Altitude
   2000 m

Safety Specifications
Overvoltage category
   Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V
Agency Approvals
   2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V

Mechanical & General Specifications
Size
   203 x 100 x 50 mm (with holster)
Weight
   624 g
Warranty
   3 years
Battery life Alkaline
Meter use:  1000 hours
Insulation test use:  Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
IP Rating
   IP40

Lebih maklumat tentang Pacific Radio (M) Sdn Bhd
Pacific Radio (M) Sdn Bhd
Pacific Radio (M) Sdn Bhd Measuring Instruments Supplier Malaysia, Soldering Irons and Switches Supplier Kuala Lumpur, KL, Audio Equipments Supplier Singapore ~ Pacific Radio (M) Sdn Bhd