MEMMERT - Universal Oven (UF750)
Brand: Memmert
Model: UF750
Product Descriptions:
The universally aplicable lab oven U is Memmert's classic appliance for temperature control in science, research and material tests in industry. The technologically perfected masterpiece made of high-quality, hygienic, easy-to-clean stainless steel leaves nothing to be desired in terms of ventilation and control technology, overtemperature protection and precisely tuned heating technology.
Product Specifications:
Temperature
Working temperature range | at least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature at +300°C |
Setting accuracy temperature | up to 99.9°C: 0.1 / from 100°C: 0.5 |
Setting temperature range | +20 to +300°C |
Temperature sensor | 1 Pt100 sensor DIN class A in 4-wire-circuit |
Language Setting | German, English, Spanish, French, Polish, Czech, Hungarian, |
Control COCKPIT | SingleDISPLAY, Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function Setpoint WAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
Adjustable Parameters | temperature (Celcius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime |
Fan | forced air circulation by 2 quiet air turbines, adjustable in 10% steps |
Fresh air | Admixture of pre-heated fresh air by electronically adjustable air flap |
Vent | vent connection with restrictor flap |
Documentation | programme stored in case of power failure |
Programming | AtmoCONTROL software for reading out, managing organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand) |
Temperature control | adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Autodiagnostic system | for fault analysis |
Internals | 2 stainless steel grid(s), electropolished |
Works calibration certificate | Calibration at 160°C |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Dimensions | w(A) x h(B) x d(C) : 1040 x 1200 x 600 mm (d less 39 mm for fan) |
Interior | easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Volume | 749 l |
Max. number of internals | 14 |
Max. loading of chamber | 300 kg |
Max. loading per internal | 30 kg |
Dimensions | w(D) x h(E) x d(F) : 1224 x 1720 x 784 mm (d +58mm door handle) |
Installation | on lockable castors |
Housing | rear zinc-plated steel |
Voltage Electrical load | 400 V and 3x 320 V w/o neutral, 50/60 Hz approx. 7000 W |
Set up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
Altitude of installation | max. 2,000 m above sea level |
Ambient temperature | +5 °C to +40 °C |
Humidity rh | max. 80%, non-condensing |
Overvoltage category | II |
Pollution degree | 2 |