Model No. (Order Code)
DT4253 |
Digital Multimeter, With mA DC, temperature |
FEATURES
Digital multimeter ideal for instrumentation, airconditioning equipment, and gas-burning devices
The DT4253, which is part of Hioki’s DT4250 series of standard digital multimeters, is ideal for measuring instrumentation, airconditioning equipment, and gas-burning devices.
・DC current: 60.00 μA / 600.0 μA / 6.000 mA / 60.00 mA
・Temperature: -40.0°C to 400.0°C
・Accepts an AC clamp sensor
600.00 μA range for measuring very low currents used by gas-burning devices
Accurately measure the very low currents such as burner flame currents.
Percentage display for instrumentation signal 4 to 20 mA measurements
The instrument can convert and display instrumentation signals of 4 to 20 mA or 0 to 20 mA representing properties such as temperature, pressure, and flow rate from a transducer or other device that converts sensor data into an electrical signal.
Clamp-on sensor based current measurement eliminates the cause of accidents
Mistakenly measuring voltage using the current range may lead to a short-circuit. The DT4253 eliminates the root cause of such accidents by providing clamp-on sensor based current measurement functionality instead of using conventional probes.
Measure output voltage on the secondary sides of inverters by eliminating harmonic components
Accurately measure the fundamental wave alone by eliminating harmonic components with the DMM’s low-pass filter function.
Measurement data can be downloaded to, reviewed on, and saved to a computer.
If you connect the DT4253 to a computer with the optional Communication Package DT4900-01, you can graph data stored in the instrument’s internal memory and display measurement results in real time on the computer. In addition, displayed measured values can be saved (in text format).
*The computer and multimeter are electrically isolated by means of optical communications so that data can be sent with peace of mind.
True RMS measurement for accurate measurement of even distorted current waveforms
Digital multimeters implement two types of measurement: mean-value measurement and true RMS measurement. The two methods yield significantly different measurement results for current waveforms, which are often distorted. For accurate measurement, it is necessary to use an instrument with true RMS capability.
Designed and manufactured in Japan to ensure high quality and guaranteed with a 3-year warranty for peace of mind
All development, design, and manufacturing processes for almost all Hioki digital multimeters are carried out at our Head Office in Nagano Prefecture. Some of the industry’s most advanced technological capabilities enable us to deliver products of the highest possible quality.
Safe measurement requires use of an instrument that suits the measurement location.
To ensure operators’ ability to use measuring instruments safely, IEC 61010 classifies the locations in which instruments are used into a series of safety-based measurement categories (ranging from CAT II to CAT IV). Using an instrument that does not satisfy the required safety level can lead to an electrical accident.
SPECIFICATIONS
Basic specifications
Accuracy guaranteed: 1 year
Regarding DMM Accuracy: Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the brochure for detailed accuracy information.
DC Voltage range |
600.0 mV to 1000 V
5 ranges, Basic accuracy: ±0.3 % rdg. ±5 dgt. |
AC Voltage range |
6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz
Basic accuracy 40 - 500 Hz : ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
AUTO AC/DCV |
Yes |
Resistance range |
600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7 % rdg. ±5 dgt. |
DC Current range |
60.00 μA to 60.00 mA, 4 ranges,
Basic accuracy: ±0.8 % rdg. ±5 dgt. |
From 4 to 20mA Percentage conversion display |
Yes |
AC Current range
(use with Clamp on probes) |
10.00 A to 1000 A, 7 ranges, Add the Clamp on probe accuracy to basic accuracy 40 - 1 kHz : ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
Temperature
(thermocouples) |
K: -40.0 to 400.0 °C , Add the Temperature probe accuracy to basic accuracy: ±0.5 % rdg. ±2 °C |
Voltage detection |
N/A |
Capacitance range |
1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9 % rdg. ±5 dgt. |
Frequency range |
99.99 Hz to 99.99 kHz, 4 ranges (limited by the minimum detectable voltage),
Basic accuracy: ±0.1 % rdg. ±1 dgt. |
Continuity check |
Continuity threshold [ON]: 25 Ω or less ,
[OFF]: 245 Ω or more,
Response time: 0.5 ms or more |
Diode test |
Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less,
Threshold of forward voltage: 0.15 V to 1.5 V |
Other functions |
Filter function, display value hold, auto hold, MAX/MIN/Average value display, relative value display, auto-power save, USB communication (option), 4-20 mA percentage conversion |
Display |
Main and Sub displays: 4-digits LCD, max. 6000 digits, bar graph |
Display refresh rates |
5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s) |
Power supply |
LR03 alkaline batteries ×4, Continuous use: 130 hours (backlight OFF) |
Dimensions and mass |
84 mm (3.31 in)W × 174 mm (6.85 in)H× 52 mm (2.05 in)D,
390 g (13.8 oz) (including batteries and holster) |
Included accessories |