Features:
1. Throughput 60T/H.
2. 14 inch touch screen.
3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting.
4. 3D holographic scattergram displays the accurate 5 part differentiation of WBC.
5. Large storage capacity: 100,000 results(including histogram, scattergram, patient information).
This 5-part auto hematology analyzer is applicable for detecting the parameters of WBC, RBC, PLT, HGB, etc. (see Section 3.2. Parameters for details) in anticoagulated venous whole blood or capillary blood, as well as WBC 5-part differential analysis and WBC counting.
Models:
Model |
BK-6310 |
|||
Throughput |
60Tests/hour |
|||
Assay Items |
5 parts, 29parameters,3 histograms, 3Dscattergram |
|||
Principle |
Tri-angle laser scatter,Flow cytometrymethod, 3D scattergram analysis, Impedancemethodfor RBC and PLT counting, Cyanide-free method for HGB test |
|||
Test Mode |
CBC mode,CBC+DIFF mode Venous whole blood, Capillary whole blood and Prediluted. |
|||
Parameters |
WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT,PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%,BAS%,NEU#, LYM#, MON#, EOS#, BAS# 4 Research parameter: ALY%, ALY#, IG%, IG# |
|||
Performance |
Item |
Linearity range |
Carry Over |
CV |
WBC |
(0~300.0x10^9)/L |
≤0.5% |
≤2.0% |
|
RBC |
(0~8.00x10^12)/L |
≤0.5% |
≤1.5% |
|
HGB |
(0~250)g/L |
≤0.5% |
≤1.5% |
|
PLT |
(0~3000x10^9)/L |
≤1.0% |
≤4.0% |
|
Sample Volume |
CBC+DIFF mode≤20μl CBC mode≤10μl |
|||
Storage |
10,000 resultsincludinghistograms,scattergrams and patient information |
|||
Interface |
4 USB ports, 1LAN port Bi-direction LIS, support HL7 protocol,Internal RFID reader |
|||
Power Supply |
AC220V±10% 50/60Hz; 110V±10% 60Hz; 400VA |
|||
Package Size |
670*590*790mm(W*D*H) |
|||
Gross Weight |
53kg |
|||
Reagent Package Size |
330*330*330mm , 400*290*240mm , 400*290*240mm |
|||
Reagent Gross Weight |
22kg, 2kg, 1kg |