TD1230 Verification Device for Withstanding Voltage Tester
Brand: Tunkia
Model: TD1230
Product Definition:
TD1230 applied to calibrate AC/DC withstanding voltage tester, withstanding voltage part of safety performance comprehensive tester, withstanding voltage part of insulation withstanding voltage tester and etc.
1. Summary
TD1230 applied to calibrate AC/DC withstanding voltage tester, withstanding voltage part of safety performance comprehensive tester, withstanding voltage part of insulation withstanding voltage tester and etc.
2. Features
• AC/DC voltage measurement:100 V~15 kV.
• AC/DC current measurement:0.1 mA~250 mA.
• Distortion / Ripple function and voltage duration time test function.
• Built-in adjustable resistive load with fineness of 1 kΩ.
• Optional high power load box (TD1240) help calibrate the output capacity.
• Oscilloscope and data statistical analysis function.
• LCD touch screen.
• Setting output values by keys.
• Notification of wire connection.
3. Applications
5. Specifications:
5.1 AC/DC Voltage Measurement
Range | Resolution |
Accuracy ±(ppm of reading + ppm of range) [1] |
|
Class 0.5 | Class 0.2 | ||
5 kV | 0.1 V | 4000 + 1000 | 1800 + 200 |
10 kV[2] | 1 V | 4000 + 1000 | 1800 + 200 |
15 kV[2] | 1 V | 4000 + 1000 | 1800 + 200 |
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). Note [2]: 10 kV, 15 kV ranges are selectable |
Range | Resolution |
Accuracy ±(ppm of reading + ppm of range) [1] |
|
Class 0.5 | Class 0.2 | ||
2 mA | 0.1 μA | 4000 + 1000 | 1800 + 200 |
20 mA | 1 μA | 4000 + 1000 | 1800 + 200 |
200 mA | 10 μA | 4000 + 1000 | 1800 + 200 |
Parameters | Range | Resolution | Accuracy |
AC Frequency | 45 Hz~65 Hz | 0.1 Hz | ± 0.2% |
AC Voltage Distortion | 0.5%~10% | 0.01% | ± 0.5% |
DC Voltage Ripple | 0.1%~30% | 0.01% | ± 0.2% |
Voltage Duration Time | 0.1 s~1000 s | 0.001 s |
AC:± (0.2%*RD + 2 periods) DC:± (0.2%*RD + 40 ms) |
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance |
Working temperature:0℃~45℃; Storage temperature:-20℃~70℃ |
Humidity Performance |
Working humidity:< 80% @ 30℃,< 70% @ 40℃,< 40% @ 50℃ Storage humidity:(20%~80%) R·H,non-condensing |
Interface | RS232 |