Model |
NR-TH08 *1 |
|||
Type |
High-accuracy temperature/voltage measurement unit |
|||
Input method |
Floating unbalanced input, channels insulated from each other, CH inputs insulated from other units (PC), |
|||
Number of channels |
Input: 8 *2 |
|||
Measurement cycle |
Fastest sampling cycle: 100 ms (A/D integration time: 2 ms) to 1 h |
|||
A/D converting system |
Delta-sigma method |
|||
A/D resolution |
24 bit |
|||
A/D integration time |
2 ms, 16.7 ms, 20 ms |
|||
Input type |
Voltage: ±50 V, ±10 V, ±5 V, ±1 V, ±0.5 V, ±0.1 V; Thermocouple; K, J, E, T, R, S, B, N, C; |
|||
Measurement range |
±50 V |
Measurable range: −55.00 to +55.00 V |
||
±10 V |
Measurable range: −11.000 to +11.000 V |
|||
±5 V |
Measurable range: −5.500 to +5.500 V |
|||
±1 V |
Measurable range: −1.1000 to +1.1000 V |
|||
±500 mV |
Measurable range: −550.0 to +550.0 mV |
|||
±100 mV |
Measurable range: −110.00 to +110.00 mV |
|||
K |
Measurable range: −100 to +1372°C -148°F to +2501.6°F, −200 to −100°C -328°F to -148°F |
|||
J |
Measurable range: −100 to +1200°C -148°F to +2192°F, −200 to −100°C -328°F to -148°F |
|||
E |
Measurable range: −100 to +1000°C -148°F to +1832°F, −200 to −100°C -328°F to -148°F |
|||
T |
Measurable range: −100 to +400°C -148°F to +752°F, −200 to −100°C -328°F to -148°F |
|||
N |
Measurable range: 0 to 1300°C 32°F to 2372°F |
|||
C |
Measurable range: 1500 to 2315°C 2732°F to 4199°F, 0 to 1500°C 32°F to 2732°F |
|||
R |
Measurable range: 300 to 1768°C 572°F to 3214.4°F, 0 to 300°C 32°F to 572°F |
|||
S |
Measurable range: 300 to 1768°C 572°F to 3214.4°F, 0 to 300°C 32°F to 572°F |
|||
B |
Measurable range: 400 to 600°C 752°F to 1112°F, 600 to 1820°C 1112°F to 3308°F |
|||
Pt100 |
Measurable range: −200 to +660°C -328°F to +1220°F |
|||
JPt100 |
Measurable range: −200 to +510°C -328°F to +950°F |
|||
Reference contact compensation |
Switchable between internal and external (individual unit setting, 0°C 32°F for external) |
|||
Reference contact compensation accuracy |
±0.7°C ±1.3°F (23°C ±2°C 73.4°F ±3.6°F, at input terminal temperature equilibrium) |
|||
Maximum input voltage |
±60 V (rated) |
|||
Input impedance |
Thermocouple, Voltage: 5 V range or lower |
10 MΩ or more |
||
Voltage: 50 V, 10 V range |
Approx. 1 MΩ |
|||
Withstand voltage |
Measurement input terminal - System bus |
1500 VAC (50/60 Hz) for 1 minute |
||
Mutually between measurement input terminals |
120 Vp-p AC/DC |
|||
Input signal source resistance |
Voltage/thermocouple |
2 kΩ or less |
||
Resistance thermometer |
10 Ω or less per line (three lines must be equal) |
|||
Thermocouple burnout |
Wire breakage detected in the thermocouple range by applying 1 μA of current |
|||
Burnout detection frequency |
Detected in the measurement cycle |
|||
Buffer memory |
200k data |
|||
Warm up time |
30 minutes or more*5 |
|||
Rating |
Power consumption |
1.2 W or less |
||
Environmental resistance |
Ambient temperature |
0 to +40°C 32°F to 104°F |
||
Relative humidity |
10 to 85% RH (no condensation) |
|||
Weight |
Approx. 230 g 8.12 oz (including approx. 100 g 3.53 oz for the OP-51582 (NR-TH08 terminal block)) |
|||
*1 The above specifications are the values when zero point adjustment is performed after the warm-up period in an environment with a temperature of 23°C ±3°C 73.4°F ±5.4°F. |