Penerangan
StrataVac Dual Vacuum Gauge | Includes bracket for pole mounting and 2 Agilent Thermocouple vacuum sensors
Measure 2 vacuum points of your system – simultaneously monitor your pump and your vacuum system
-
Measure your process in the .001 to 760 Torr range | thermocouple sensors best accuracy is between .001 – 6 Torr
-
Range and accuracy for sensors:
-
Total Range: 1 millitorr to 760 Torr
-
Range with Accuracy: Agilent 531 and 536 Sensors: .001 Torr to 6 Torr
-
For rough vacuum measurement, we recommend StrataVac 775i with Piezo sensor for measurement and accuracy from 1 Torr to 775 Torr
-
Calibrated—Every STRATAVAC vacuum instrument is calibrated and pretested under actual vacuum against a NIST standard. If you need NIST certification with traceable data included, add this option to your order
-
Applications— Vacuum Ovens, Vacuum Furnaces, Vacuum Bagging / Composites, Rotary Evaporation, Laboratory, R&D, Foreline monitoring, distillations, Schlenk line monitoring
This STRATAVAC Dual Vacuum Gauge Package Includes:
-
StrataVac vacuum gauge controller with bracket for pole mounting
-
Two (2) Sensor Cards to drive the sensors
-
TWO (2) Agilent Thermocouple Sensors
-
Choose either the Agilent 531 NPT sensor or the Agilent 536 sensor which is available in different vacuum installation interfaces (NW/KF)
-
The Agilent 536 sensor is a welded 304 stainless steel version of the 531 with an additional baffle that greatly extends the sensor life. (choose either KF25, KF16, or NPT)
3 Simple Steps for Vacuum Process Set-Up with STRATAVAC:
-
Plumb the vacuum gauge(s) and valve into your system
-
Connect the sensor, power, and I/O cables into STRATAVAC
-
Measure vacuum
Note: this set-up is for vacuum measurement only. Want to control with upper and lower set-points? Contact us for a quote that includes relays. If you need vacuum control then choose a StrataVac control bundle with valves
here
StrataVac Software: Connect to StrataCapture software by USB port on StrataVac to your laptop.
-
Data Graphing and logging: through USB serial using Stratacapture software to display, graph and log vacuum data
-
Watch this video of DigiVac’s CEO, Tim Collins, demonstrating how vacuumnetwork.org and StrataCapture (StrataVac’s companion software)
Technical Specifications
Units |
Torr, mbar, kPa |
Control Range |
.001 to 760 Torr |
Range with Accuracy |
Dependent on sensor |
Sensor Type |
536 KF25, 536 KF15, 536 NPT, or 531 NPT |
Dimensions |
1.7”H x 3.52”W x 5.35” Deep |
Certifications |
– |
Lebih maklumat tentang Hexo Industries (M) Sdn Bhd