HIOKI DSM-8104 Digital Super Megohmmeter

HIOKI DSM-8104 Digital Super Megohmmeter

Kategori: Super Megohmmeters, Electrometers, Picoammeters Tersedia
Untuk maklumat lanjut, lawati laman web rasmi kami di marktest.com.my

Penerangan

HIOKI DSM-8104 Digital Super Megohmmeter
0.1V to 1000V/1x10^3Ω to 3x10^16Ω Digital Super Megohmmeter for Testing Leak Current of Capacitive or Highly Insulated Materials
 
DiscontinuedHIOKI SM7110 Super Megohm Meter
 

OVERVIEW
  
The Hioki DSM Series Super Megohmmeters are capacitance and high insulation resistance meters with a 0.1 fA resolution for testing current. The DSM-8104 is a single-channel meter, ideal for practical measurement applications in semiconductor and electronic materials research.

Key Features

Model No. (Order Code)
DSM-8104 Discontinued, Digital Super Megohmmeter, 1 ch

Note: RS-232C connection cable: The optional RS-232C Cable 9637 cannot be used with this product. Use a cross cable for connections as described in the instruction manual.
 

SPECIFICATIONS

Basic specifications

Accuracy guaranteed: 1 year
 
Number of channels 1 ch
DC current measurement 10 pA range (0.1 fA resolution), Accuracy: ±(3.0 % of rdg. +1.2 % of range)
100 pA range (1.0 fA resolution), Accuracy: ±(1.5 % of rdg. +0.6 % of range)
1 nA range (10 fA resolution), Accuracy: ±(0.6% of rdg. +0.6% of range)
10 nA range (100 fA resolution), Accuracy: ±(0.4 % of rdg. +0.5 % of range)
100 nA range (1 pA resolution), Accuracy: ±(0.4 % of rdg. +0.5 % of range)
1 μA range (10 pA resolution), Accuracy: ±(0.4 % of rdg. +0.5 % of range)
10 μA range (100 pA resolution), Accuracy: ±(0.4 % of rdg. +0.5 % of range)
100 μA range (1 nA resolution), Accuracy: ±(0.4 % of rdg. +0.5 % of range)
(1) With 300 ms measurement time, with averaging processing enabled
(2) At a temperature of 23 °C ±5 °C with humidity of 85% rh
(3) After self-calibration (1 min. interval)
(4) With averaging processing enabled
Resistance measurement capabilities 1 × 10^3 Ω to 3 × 10^16 Ω (Open-circuit), (Measurement time is 300 ms)
Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy.
Setting voltage range
(Accuracy)
0.1 to 250.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±150 mV
251 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±400 mV
Current limiter 0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA
Measurement time setting Delay: 0 to 9,999 msec, Sampling time: 2 to 300 msec
Functions Comparator measurement, deviation measurement, percentage measurement, surface resistivity, volume resistivity, voltage monitor, contact check
Program function 10 types of discharge, charge, measure and measurement sequence discharge patterns can be programmed.
Data storage Up to 1,000 measurement values can be stored
Display LCD (8 lines of 30 characters), with backlight, High voltage warning indicator
Interfaces GP-IB, RS-232C, Handler Interface
Power supply 100 V AC ±10% (standard), (115, 220 or 240 V AC factory option), 50/60 Hz, 55 VA
Dimensions and mass 332 mm (13.07 in) W × 89 mm (3.50 in) H × 450 mm (17.72 in) D, 6.7 kg (236.3 oz)
Included accessories Power cord ×1, Instruction manual ×1