Detection mode | MEAN/RMS (switch) |
---|---|
DC V |
50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V (Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V]) ±0.02%rdg±2dgt (Basic accuracy) |
AC V [RMS] |
50.000*/500.00mV/5.0000/50.000/500.00/1000.0V(* Only for 1062) (Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V]) ±0.4%rdg±30dgt (Basic accuracy) |
AC V [MEAN] |
50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V (Input impedance:11MΩ<50pF [50/500mV/5V],10MΩ<50pF [50/500/1000V]) ±1%rdg±30dgt (Basic accuracy) |
DCV+ACV |
5.0000/50.000/500.00/1000.0V (Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V]) ±0.5%rdg±10dgt (Basic accuracy) |
DC A |
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A ±0.2%rdg±5dgt (Basic accuracy) |
AC A [RMS] |
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A ±0.75%rdg±20dgt (Basic accuracy) |
AC A [MEAN] |
500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A ±1.5%rdg±20dgt (Basic accuracy) |
DCA+ACA |
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A ±1%rdg±10dgt (Basic accuracy) |
Ω |
500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ ±0.05%rdg±2dgt (Basic accuracy) |
LowPower-Ω |
5.000/50.00/500.0kΩ/5.000MΩ ±0.2%rdg±3dgt (Basic accuracy) |
Continuity buzzer | 500.0Ω (The buzzer turns on for resistances lower than 100±50Ω) |
Diode test |
2.4000V ±1%rdg±2dgt Open curcuit voltage: <5V (Approx. 0.5mA Measuring Current) |
Capacitance |
5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF ±1%rdg±5dgt (Basic accuracy) |
Frequency |
2.000 - 9.999/9.00 - 99.99/90.0 - 999.9Hz/0.900 - 9.999/9.00 - 99.99kHz ±0.02%rdg±1dgt (Basic accuracy) |
DUTY | 10 - 90% ±1%rdg |
Temperature |
-200 - 1372ºC ±1%rdg±1.5ºC -328 - 2501.6ºF ±1%rdg±2.0ºF (with the use of K-type Temperature probe) |
Measurement functions | DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω * |
Other functions | Data Hold (D/H), Auto Hold (A/H), Peak Hold * (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight |
Applicable standards |
IEC 61010-1 CAT IV 600V, CAT III 1000V Pollution degree 2, IEC 61010-031, IEC 61326-1 (EMC) |
Power source | R6 (1.5V)×4 (Auto power off: approx. 20 minutes) |
Dimensions | 192(L)×90(W)×49(D) mm |
Weight | Approx. 560g (including batteries) |