Rhopoint - Novo-Gloss Trio Glossmeter

Rhopoint - Novo-Gloss Trio Glossmeter

Category: Gloss/DOI/Haze/Appearance Testers Available
For more information, visit our official website at obsnapinstrument.com

Description

The Novo-Gloss Trio low cost glossmeter is ideal for gloss measurements of all surfaces and is suitable for matt to mirror gloss surfaces

60° is referred to as the universal measurement angle and is the most commonly specified geometry in applications such as paints, coatings, plastics, automotive interiors and general manufacturing. It can be used as a basic gloss assessment for any surface

20° is the angle selected for high gloss applications such as paints and coatings, polished metals as it gives an improved resolution for high gloss finishes.

Advanced features including onboard statistics and the ability to download readings, the 20/60/85° glossmeter is perfect for laboratory applications

Why Measure Gloss?

Gloss is an aspect of the visual perception of objects that is as important as colour when considering the psychological impact of products on a consumer.

It has been defined as 'The attribute of surfaces that causes them to have shiny or lustrous, metallic appearance.'The gloss of a surface can be greatly influenced by a number of factors, for example the smoothness achieved during polishing, the amount and type of coating applied or the quality of the substrate.

Manufacturers design their products to have maximum appeal: highly reflective car body panels, gloss magazine covers or satin black designer furniture.

d5c776458be2211f966676ac4742282f

It is important therefore that gloss levels are achieved consistently on every product or across different batches of products.

Gloss can also be a measure of the quality of the surface, for instance a drop in the gloss of a coated surface may indicate problems with its cure, leading to other failures such as poor adhesion or lack of protection for the coated surface.

It is for these reasons that many manufacturing industries monitor the gloss of their products, from cars, printing and furniture to food, pharmaceuticals and consumer electronics.

How is Gloss Measured?

Gloss is measured with a gloss-meter shining a known amount of light at a surface and quantifying the reflectance. The angle of the light and the method by which the reflectance is measured are determined by surface and also aspect of the surface appearance to be measured.
how to measure gloss with a gloss meter

Which Angle Should I Use For My Application?

ISO 2813 and ASTM D523 (the most commonly used standards) describe three measurement angles to measure gloss across all
surfaces.

Gloss is measured in gloss units (GU) and is traceable to reference standards held at BAM (Germany), NRC (Canada) or NPL (UK).

Universal Measurement Angle60°

All gloss levels can be measured using the standard measurement angle of 60°. This is used as the reference angle with the complimentary angles of 85° and 20° often used for low and high gloss levels respectively.

Low Gloss85°

For improved resolution of low gloss a grazing angle of 85° is used to measure the surface. This angle is recommended for surfaces which measure less than 10GU when measured at 60º.

This angle also has a larger measurement spot which will average out differences in the gloss of textured or slightly uneven surfaces.

High Gloss20°

The acute measurement angle of 20° gives improved resolution for high gloss surfaces. Surfaces that measure 70GU and above at the standard angle of 60° are often measured with this geometry.

The 20° angle is more sensitive to haze effects that affect the appearance of a surface.

To quantify haze, distinctness of image, reflected image quality and other surface texturing please consider the Rhopoint IQ.

 

Novo-Gloss Trio: low cost glossmeter measurement
Fast measurement of all parameters. Full on-board statistics with trend analysis and reporting.
trio low cost glossmeter main screen
Simultaneous measurement of all parameters, results are and time stamped.
trio glossmeter statistics
Displays full statistics for the number of readings in the current batch.
trio glossmeter graph
Graphical reporting for quick trend analysis.
trio glossmeter pass-fail
Pass / fail parameters can be defined for instant identification of non-conformances.

Calibration
The instrument detects when the calibration tile is attached to the instrument enabling faster and easier calibration
Novo-Gloss gloss meter calibration tile detection
Rapid data transfer0cdeac4f93df8211903b410f7bcf2828
Software-free data transferUSB connection to PC instantly recognises the device as a drive location which facilitates the quick transfer of files using Windows Explorer or similar.Direct data input via Bluetooth892122db32fac78ac96bdb8738922dac
Instantly transmit measured readings directly to programs such as MS Excel on your PC / tablet / smartphone to greatly simplify the reporting process.

175c8375a232c575cc2bf5be5022c5c8

Power3c62e8ffebf9dc993eed4f0591cadce7
10,000+ readings per charge

The instrument is rechargeable via USB/PC or mains.

Gloss

60°, 20/60°* and 20/60/85°* versions available

20° Improved accuracy and resolution on high gloss & metallic samples (>70GU when measured at 60°)
60° Universal angle – all gloss levels
85° Improved resolution for matt finishes 

Resolution 0.1GU • Repeatability ±0.2GU • Reproducibility ±0.5GU (0-100GU)
Resolution 0.1GU • Repeatability ±0.2% • Reproducibility ±0.5%(100-2000GU)

Measurement range:
20º: 0-2000GU
60º: 0-1000GU
85º: 0-199GU

Standards:
ISO 2813, ISO 7668
ASTM D523, ASTM D2457
DIN 67530
JIS 8741, JIS K 5600-4-7

Instrument Specification

Operation

  • Full colour easy to read screen
  • Adjustable brightness
  • 6 button touch sensitive interface

Construction

  • Integrated calibration holder with in-position detector for
    error free calibration

Measurement

  • Single button push to measure all parameters
  • Fast measurement
  • Results batching with user definable names

Statistical Analysis

  • Max, min, mean, S.D.

Graphical Analysis

  • On board trend analysis
  • Gloss values

Power

  • Rechargeable lithium ion
  • 17+ hours operation
  • 10,000+ readings per charge

Memory

  • 8MB= >2000 readings
  • User defineable alphanumeric batching

Data Transfer

  • Bluetooth
  • PC compatible
  • USB connection, no software install required

Measurement Area

  • 20°: 6mm x 6.4mm
  • 60°: 6mm x 12mm
  • 85°: 4.4mm x 44mm
  • Operating Temperature: 15 - 40 ° C (60 - 104 ° F)
  • Humidity: Up to 85%, non condensing

Dimensions & Weight

  • 65mm x 140mm x 50mm (H x W x D)
  • 20/60/85º = 390g
  • Packed weight: 1.6kg
  • Packed dimensions: 110mm x 280mm x 220mm (H x W x D)
  • Commodity code: 9027 5000

More detail about Obsnap Instruments Sdn Bhd
Obsnap Instruments Sdn Bhd
Obsnap Instruments Sdn Bhd Testing Instruments Supplier Malaysia, Measuring Equipments Supply Selangor, Simulation Systems Supplies Kuala Lumpur (KL) ~ Obsnap Instruments Sdn Bhd
Contact Us flagMalaysia