Description
General Details
RESmart BPAP Series provide non-invasive ventilation for patients with respiratory insufficiency in the hospital or at home. The breath by breath pressure relief of Reslex offers most natural respiratory therapy. It can also offer the ultimate in sleep therapy comfort, with auto adjusting pressure delivery.
Reliable stability
-
Monitored parameters include: Expiratory tidal volume (Vte), Respiratory rate(Res. Rate),Minute ventilation (Min Vent), Leakage (Leak)
-
Quick and easy settings when in use, including: IPAP, EPAP, Res.Rate, E/I Rate,I sens, E sens, Rise time.
-
Innovative tracking technology ensures accurate and comfortable therapy.
-
Unique sensitivity setting offers individual therapy algorithm, makes every userget the most comfort.
-
Auto on/Auto off.
-
Alert when accidental power off.
-
In time alert function when mask/tubing off line.
-
Automatic leakage and altitude compensation.
Ergonomic design
-
Integrated and knock-down designed InH2 heated humidifier.
-
DC 24V powered and infrared controlled humidifier make user safe andcomfortable.
-
Patented anti-countercurrent water tank.
-
Unique Delay-off feature protects RESmart BPAP from humidity hazard.
-
Powerful embedded flash memory and SD card stores more than 30 nights of fullraw airflow data and 365 nights of user blog.
Techical Info
Device Size
Dimensions: |
220 × 194 × 112 mm
|
313 × 194 × 112 mm (with InH2 heated humidifier)
|
Weight: |
< 2.2 kg
|
< 3 kg (with InH2 heated humidifier)
|
Product Use, Transport and Storage
|
Operation |
Transport and Storage
|
Temperature: |
5 to 35 °C |
-20 to 55 °C
|
Humidity: |
≤80% Non-condensing |
≤93% Non-condensing
|
Atmospheric Pressure: |
860 to 1060 hPa |
500 to 1060 hPa
|
AC Power Consumption
100 – 240 V AC, 50/60 Hz, Max 95 VA
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IPX1 – Drip-Proof, Vertical
Mode of Operation:
Continuous
Model Comparison
Product Model |
Pressure |
Work Mode
|
RESmart BPAP 20S |
4-20 cm H2O |
CPAP, S
|
RESmart BPAP 20A |
4-20 cm H2O |
CPAP, S, Auto S
|
RESmart BPAP 20T |
4-20 cm H2O |
CPAP, S, T, S/T
|
RESmart BPAP 25 |
4-25 cm H2O |
CPAP, S
|
RESmart BPAP 25A |
4-25 cm H2O |
CPAP, S, Auto S
|
RESmart BPAP 25T |
4-25 cm H2O |
CPAP, S, T, S/T
|
RESmart BPAP 30T |
4-30 cm H2O |
CPAP, S, T, S/T
|
SD card capacity
2 G bytes
Pressure Range
For 20S/ 20A/ 20T:
IPAP: 4 cm H2O to 20 cm H2O (in 0.5 cm H2O increments)
EPAP: 4 cm H2O to 20 cm H2O (in 0.5 cm H2O increments)
For 25/ 25A/ 25T:
IPAP: 4 cm H2O to 25 cm H2O (in 0.5 cm H2O increments)
EPAP: 4 cm H2O to 25 cm H2O (in 0.5 cm H2O increments)
For 30T:
IPAP: 4 cm H2O to 30 cm H2O (in 0.5 cm H2O increments)
EPAP: 4 cm H2O to 25 cm H2O (in 0.5 cm H2O increments)
Pressure Stability
For 20S/ 20A/ 20T:
IPAP: 4 cm H2O to 20 cm H2O (in 0.5 cm H2O increments)
EPAP: 4 cm H2O to 20 cm H2O (in 0.5 cm H2O increments)
For 25/ 25A/ 25T:
Static Pressure Stability: 4 to 25 cm H2O (± 0.5 cm H2O)
Dynamic Pressure Stability: 4 to 25 cm H2O (± 1.0 cm H2O)
For 30T:
Static Pressure Stability: 4 to 30 cm H2O (± 0.5 cm H2O)
Dynamic Pressure Stability: 4 to 30 cm H2O (± 1.0 cm H2O)
Altitude Compensation
Automatic altitude compensation
Manually setting, level 0, 1, 2
Sound Pressure Level
< 30 dB, when the device is working at the pressure of 10 cm H2O
Maximum Flow
For 20S/ 20A/ 20T:
Test Pressure (cm H2O) |
4 |
8 |
12 |
16 |
20 |
Average flow at the patient connection port (L/min) For 25/ 25A/ 25T: |
70.1 |
72.1 |
72.3 |
70.1 |
73.4
|
Test Pressure (cm H2O) |
4 |
10 |
15 |
20 |
25 |
Average flow at the patient connection port (L/min) For 30T: |
70.1 |
73.0 |
64.0 |
73.4 |
74.6
|
Test Pressure (cm H2O) |
4 |
10.5 |
17 |
23.5 |
30
|
Average flow at the patient connection port (L/min) |
70.1 |
73.0 |
70.5 |
71.7 |
72.8 |