HIOKI RESISTANCE METER RM3542A

 

High-Speed Resistance Meter Ideal for Automated Lines; Compatible with Super-Small Electronic Components


Hioki DC resistance meters measure a broad range of resistance values at a high level of precision. The RM3542A delivers high-speed and high-stability measurements and improved productivity by minimizing variations and reducing contact errors. In addition, the low-impact measurement technology lets you measure resistance for miniature 008004-size electronic parts.


Key Features
• Applied voltage limit function lets you switch the detection voltage to 5 V or less
• Contact improvement function suppresses rush current to aid in probing of super-small components
• Extensive selection of measurement ranges ensures the right detection voltage and delivers stable measurement
• Scaling function corrects for mounting state and test stage differences

 
RM3542-50  
RM3542-51 Built in GP-IB interface

Test fixtures are not supplied with the unit. Please select an optional test fixture when ordering.

 

Guarantee3years
 
GPIB
RM3542-51
 
RS232C
 
CE
 
Resistance range [at Low Power OFF] 100 mΩ range (Max. 120.0000 mΩ, 0.1 μΩ resolution) to 100 MΩ range (Max. 120.0000 MΩ, 100 Ω resolution), 16 steps
[at Low Power ON] 1000 mΩ range (Max. 1200.000 mΩ, 1 μΩ resolution) to 1000 Ω range (Max. 1200.000 Ω, 1 mΩ resolution), 6 steps
Display Monochrome graphic LCD 240 × 64 dot, white LED backlight
Measurement accuracy [with SLOW mode, at 100 mΩ range] ±0.015 % rdg. ±0.002 % f.s.
[with SLOW mode, at 1000 Ω range] ±0.006 % rdg. ±0.001 % f.s. (best case)
Testing current [at 100 mΩ range] 100 mA DC to [at 100 MΩ range] 100 nA DC
Open-terminal voltage 20 V DC max. (with applied voltage limit function enabled: max. 10 V DC)
Sampling rate FAST, MEDIUM, SLOW, 3 settings
Measurement times [at 100 Ω /1000 Ω ranges, with Low Power OFF]
FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time)
Integration time 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz
Note: PLC = one power line cycle (mains wave-form period)
Other functions Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), Applied Voltage Limit Function, Scaling Function, OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,.
Interfaces RS-232C, Printer, GP-IB (Model RM3542-51)
External I/O Trigger, Hold input, Comparator output, Settings monitor terminal
Power supply 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions and mass 260 mm (10.24 in) W × 88 mm (3.46 in) H × 300 mm (11.81 in) D, 2.9 kg (102.3 oz)
Accessories Power cord ×1, EXT. I/O male connector ×1, Instruction manual ×1, Operation guide ×1