UNIPULSE – Digital Indicator with Fast Peak Hold Function
Brand: Unipulse
Model: F325
Specifications:
Analog | Excitation voltage | DC10V or 2.5V±10% (switchable) Output current: within 30mA | |||
Signal input range | -3.0 to +3.0mV/V | ||||
Accuracy Non-linearity | 0.02%/FS±1 digit (AT 3mV/V input), Zero drift: within 0.5μV/℃ RTI, Gain drift: within 25ppm/℃ | ||||
A/D converter: | Speed:30,300,3000 times/ sec (selectable), Resolution: 24 bit | ||||
Analog filter | Primary low-pass filter(select frequency from 10,30,100,300,1k,3k,10k, 30Hz | ||||
Digital filter |
Filter1: Bessel low-pass filter (secondary) Cut-off frequency can be set to any value from 1/300 to 1/10 of sampling speed. It can be turned off. 3000timmes/sec: 10 to 300Hz, 300times/sec: 1.0 to 30.0Hz, 30times/sec: 0.1 to 3.00Hz Filter2: Moving average You can set the value anywhere between 1 to 999 times |
||||
Resolution | 1/30000(at 3.0mV/V input) | ||||
Monitor output | Sensor input of approx. 2V per 1mV/V (Input resistance over 2k Ω) | ||||
Hold function | Sample or peak hold (You can chose either analogue or digital peak hold from setting menu. Frequency response of analogue hold is 1kHz | ||||
Display | Display unit | 15mm character height | : 7-segment green LED (5-digits) | ||
Numerical value | :5-digits Indicated value: -19999 to +9999 | ||||
Decimal points | :Selectable decimal point displaying position (0.000, 0.00, 0.0, 0) | ||||
Display item |
Status display Green |
: Red 3 Ø LED: HI, LO, PEAK, HOLD : 3 ØLED: OK |
|||
Display frequency | : Selectable from3,6,13 and 25 times/sec | ||||
External input | DZ, HOLD, H.RESET | ||||
(3) |
< Volt - free contact input > External devices can be connected, such as relays, switches, transistors and so on. The signal can be outputted by making short-circuit between input and the common terminal. In case the transistor is connected, please use NPN (sink) transistor Internal power supply voltage: DC12V, When short-circuited: Approx. 4mA < DC - Input for both plus common and minus common type > (please specify the type of input when ordering) External devices can be connected, such as relays, switches, transistors and so on. The input signal is outputted by applying a voltage between input and the common terminal. In case a transistor is connected, please use NPN type for plus common and PNP type for minus common. Maximum Rated Voltage: DC27. 6V, ''ON '': when the voltage is above DC 9V (Load Current: approx . 10mA at DC24V ), '' OFF '' below DC3V. |
||||
External output | HI, OK, LO, Output Selection 1, Output Selection 2 | ||||
(5) | Assign output from the various selection: HH, LL, Overload, RUN, HOLD, NZ, DZ response) PhotoMOS relay output (common for sink and source type) Maximum rated voltage: DC307, maximum rated current: 100mA, operating time: approx . 1 msec | ||||
General Specifications | Power supply voltage |
AC spec: AC100 to 40V (+10%, -15%) [free power supply 50/60Hz] DC spec: DC12 to 24V (±15%)(please specify the type of input when ordering) |
|||
Power consumption | AC spec. 3W typ/DC spec. 4W typ | ||||
Rush current (Typ) AC spec |
: 2A, 1msec:AC100V average load condition ( cold start at room temperature) : 4A, 1msec:AC200V average load condition ( cold start at room temperature) |
||||
DC spec |
: 2A, 20msec:DC12V average load condition ( cold start at room temperature) : 1A, 50msec:DC24V average load condition ( cold start at room temperature) |
||||
Operation condition |
Operation tem.range: -10 to 40℃ Humidity: 85% RH or less (non-condensing) |
Storage tem.range: -40 to +80℃ | |||
External dimensions | 96(W) X 48(H) X 132.5(D)mm (not including projections) | ||||
Weight | Approx. 600g | ||||
Attachments |
AC input cord (AC power supply is selected)..1, 3p-2p conversion adapter (AC power supply is selected..1, Terminator (RS-485 is selected)..1, BCD output connector (BCD output option is selected)..1, Operational manual..1 | ||||
Optional Accessories |
CA325AC3P-B3 CA325AC3P-CEE7/7-B2 CN3P-2P CN51 CN82 CN83 CN84 TSU03 |
: AC Supply cord 3m : AC Supply cord (Voltage resistance: 250V)2M : 3P-2P converter plug for AC input cord : BCD output connector : I/O connector (wire entry holes: front) : I/O connector (wire entry holes: bottom) : I/O connector (wire entry holes: top) : DC lighting surge unit |
|||
CE Marking Certificate |
|
|