UNIPULSE – Digital Indicator with Fast Peak Hold Function

UNIPULSE – Digital Indicator with Fast Peak Hold Function

Category: Weighing Available
For more information, visit our official website at carlssoon-tech.com

Description

UNIPULSE – Digital Indicator with Fast Peak Hold Function
Brand: Unipulse
Model: F325
 
Specifications:


Analog Excitation voltage DC10V or 2.5V±10% (switchable) Output current: within 30mA
Signal input range -3.0 to +3.0mV/V
Accuracy Non-linearity 0.02%/FS±1 digit (AT 3mV/V input), Zero drift: within 0.5μV/℃ RTI, Gain drift: within 25ppm/℃
A/D converter: Speed:30,300,3000 times/ sec (selectable), Resolution: 24 bit
Analog filter Primary low-pass filter(select frequency from 10,30,100,300,1k,3k,10k, 30Hz
Digital filter Filter1: Bessel low-pass filter (secondary)
Cut-off frequency can be set to any value from 1/300 to 1/10 of sampling speed. It can be turned off. 3000timmes/sec: 10 to 300Hz, 300times/sec: 1.0 to 30.0Hz, 30times/sec: 0.1 to 3.00Hz Filter2: Moving average
You can set the value anywhere between 1 to 999 times
Resolution 1/30000(at 3.0mV/V input)
Monitor output Sensor input of approx. 2V per 1mV/V (Input resistance over 2k Ω)
Hold function Sample or peak hold (You can chose either analogue or digital peak hold from setting menu. Frequency response of analogue hold is 1kHz
Display Display unit 15mm character height : 7-segment green LED (5-digits)
  Numerical value :5-digits Indicated value: -19999 to +9999
  Decimal points :Selectable decimal point displaying  position (0.000, 0.00, 0.0, 0)
Display item Status display
Green
: Red 3 Ø LED: HI, LO, PEAK, HOLD
: 3 ØLED: OK
  Display frequency : Selectable from3,6,13 and 25 times/sec
External input DZ, HOLD, H.RESET
(3) < Volt - free contact input > External devices can be connected, such as relays, switches, transistors and so on. The signal can be outputted by making short-circuit between input and the common terminal. In case the transistor is connected, please use NPN (sink) transistor Internal power supply voltage: DC12V, When short-circuited: Approx. 4mA
< DC - Input for both plus common and minus common type > (please specify the type of input when ordering) External devices can be connected, such as relays, switches, transistors and so on. The input signal is outputted by applying a voltage between input and the common terminal. In case a transistor is connected, please use NPN type for plus common and PNP type for minus common. Maximum Rated Voltage: DC27. 6V, ''ON '': when the voltage is above DC 9V (Load Current: approx . 10mA at DC24V ), '' OFF '' below DC3V.
External output HI, OK, LO, Output Selection 1, Output Selection 2
(5) Assign output from the various selection: HH, LL, Overload, RUN, HOLD, NZ, DZ response) PhotoMOS relay output (common for sink and source type) Maximum rated voltage: DC307, maximum rated current: 100mA, operating time: approx . 1 msec 
General Specifications Power supply voltage AC spec: AC100 to 40V (+10%, -15%) [free power supply 50/60Hz]
DC spec: DC12 to 24V (±15%)(please specify the type of input when ordering)
Power consumption AC spec. 3W typ/DC spec. 4W typ
Rush current (Typ)
AC spec
 
: 2A, 1msec:AC100V average load condition ( cold start at room  temperature)
: 4A, 1msec:AC200V average load condition ( cold start at room  temperature)
DC spec : 2A, 20msec:DC12V average load condition ( cold start at room  temperature)
: 1A, 50msec:DC24V average load condition ( cold start at room  temperature)
  Operation condition Operation tem.range: -10 to 40℃
Humidity: 85% RH or less (non-condensing)
Storage tem.range: -40 to +80℃
External dimensions 96(W) X 48(H) X 132.5(D)mm (not including projections)
Weight Approx. 600g
Attachments
 
AC input cord (AC power supply is selected)..1, 3p-2p conversion adapter (AC power supply is selected..1, Terminator (RS-485 is selected)..1, BCD output connector (BCD output option is selected)..1, Operational manual..1
Optional Accessories CA325AC3P-B3
CA325AC3P-CEE7/7-B2
CN3P-2P
CN51
CN82
CN83
CN84
TSU03
: AC Supply cord 3m
: AC Supply cord (Voltage resistance: 250V)2M
: 3P-2P converter plug for AC input cord
: BCD output connector
: I/O connector (wire entry holes: front)
: I/O connector (wire entry holes: bottom)
: I/O connector (wire entry holes: top)
: DC lighting surge unit
CE Marking Certificate
  • EMC directive EN61326-1
  • Safety standard EN61010-1, EN62311

More detail about Carlssoon Technologies (Malaysia) Sdn Bhd
Carlssoon Technologies (Malaysia) Sdn Bhd
Carlssoon Technologies (Malaysia) Sdn Bhd Testing Equipment Supplier Melaka, Measuring Instrument Supply Malaysia, Lab Supplies Ayer Keroh ~ Carlssoon Technologies (Malaysia) Sdn Bhd
Contact Us flagMalaysia