Testing voltage | 0 to 2.5 kV / 0 to 5.0 kV AC, 2 range configuration 500 VA (30 minutes rated) |
---|---|
Voltage setting | Manual setting |
Waveform/Frequency | Same as the power supply waveform, synchronized with the power supply |
Current measurement | 0.01 mA to 120 mA, True RMS rectified (digital display) |
Voltage meter | Accuracy: ±1.5 % f.s. (digital), ±5 % f.s. (analog, f.s.=5 kV) |
Current measurement | 0.01 mA to 120 mA, (Average value rectified, effective value digital display) |
Measurement range | 2 mA/8 mA (0.01 mA resolution), 32 mA (0.1 mA resolution), 120 mA (1 mA resolution) |
Voltage meter | Digital, Accuracy: ±1.5 % rdg. (f.s.=5.00 kV) |
Judgment function | Window comparator method (Digital setting) |
[Insulation test section] | |
Testing voltage | 500 V DC, 1000 V DC, Unloaded voltage: 1 to 1.2 times rated voltage |
Rated testing current | 1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V) |
Measurement range, Accuracy |
0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg., 1000 MΩ to 2000 MΩ: ±8 % rdg. |
Judgment function | Window comparator method (Digital setting) |
[Timer section] | |
Setting range | 0.5 to 999 s |
[General section] | |
Monitor function | Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s |
Power supply | 220 V AC, (50/60 Hz), 800 VA max. |
Dimensions and mass |
320 mm (12.60 in)W × 155 mm (6.10 in)H × 330 mm (12.99 in)D, 21.5 kg (758.4 oz) |
Accessories | H.V. Test lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1 |