Description
Automatic dialyzer Reprocessor
-
Fully automatic reprocessing and disinfecting cycles
-
Built-in anti-splash shield
-
Chemical detection sensor
-
Dialyzerdis infectant fill sensor
-
Membrane pressure leak test
-
Chemical concentration sensor
Technical Spec
-
Power: 100-130 VAC220-240VAC
-
Consumption: 150W
-
Dimension: H420 x W380x D420mm
Features and Benefits
Reliable and Safe
-
Fully automatic reprocessing and disinfecting cycles
-
Built-in effluent basin
-
Membrane pressure leak test
-
Low water inlet pressure alarm
-
Chemical detection sensor
-
Dialyzer disinfectant fill sensor
-
Chemical concentration sensor
User-Friendly
-
Easy to operate LCD touch screen
-
Select able or presets of dialyzer cell volume selections
-
Total cell volume measurement and display
-
Reverse ultra-filtration cleaning
-
Selectable programs for all types of dialyzer Innovative Design
-
Economical consumption of disinfectant
-
Accepts most common disinfectants
Installation
Environmental Limits
Ambient temperature: 18-35º Celsius
Relative humidity: up to 80%
Water requirements
Water quality: must comply to AAMI RD62
Water pressure: 20-50 PSI
Water flow: min. 1.5 L/min
Temperature: 20-35º Celsius
Consumption: ≈18 L/cycle
Drain requirements
Flow: min. 3.5L/min
Height: min. 200mm below TEC11
Technical Specifications
Dimensions: Height: 420mm
Width: 380mm
Depth: 420mm
Weight: ~30kg
Power: 100 –130 VAC, 60Hz220 –240 VAC, 50Hz
Consumption: 150W
Performance
Dialyzer Testing
Pressure Leak Test:
Pressure tested to 300mmHg ±10mmHg
Acceptable leakage rate is 10%
Volume Testing:
Range 10ml to 500ml
Accuracy ±5%
Reprocessing Time
Approximately 6 -12minutes per dialyzer
Disinfectants
Company recommended disinfectants
Alarms
Audio / Visual indication of:
• Low water inlet pressure
• Low disinfectant
• Disinfectant concentration (high & low)
• Leak test failure
• Total cell volume high & low alarm preset
• Cycle complete message
• Volume measurement error
More detail about DIROSYS SOLUTIONS (M) SDN BHD