MAS-100 NT Ex Industrial Microbial Air Sampler (ATEx)

MAS-100 NT Ex Industrial Microbial Air Sampler (ATEx)

Category: Microbial Air Sampler Available
Purchasable on official site
For more information, visit our official website at muser.com.my

Description

Merck 109194 MAS-100 NT Ex

Portable, battery operated instrument for microbial monitoring of ambient air in explosion hazard areas

 

The MAS-100 NT® Ex shares all of the same functions of the MAS-100 NT® system but is specially designed for use in explosion proof areas. The MAS-100 NT® Ex has received ATEx Conformity and can be used in zone 2 and gas groups 11A,11B, and 11C in temperature classes T1 to T4. 

MAS-100NT® Ex portable microbial air samplers are the industry standard for use in critical environments. These compact yet sophisticated devices are the preferred choice for those demanding the highest quality in microbial air monitoring.

The MAS-100NT® Ex systems feature a new 300-hole perforated lid for increased collection efficiency and impaction speed. Both systems utilize standard 90 – 100 mm agar plates or can be adapted to fit 55 – 60 mm contact plates allowing for a low consumable cost and greater flexibility.

Sampling at 100 litres per minute, these systems have the highest airflow accuracy available at ± 2.5 %, compared to others that can be as high as ±10 %. The integrated flow sensor allows the user to freely interchange the perforated lids without affecting the accuracy or the calibration of the unit. Sampling volumes are also easily configurable between 1 and 2,000 litres.

The units have an impact velocity of 19.6 meters per second equivalent to Anderson 6 and isokinetic flow rate that will not produce turbulence in a laminar flow environment. The new SQS function will allow for smaller sampling volumes over longer periods of time, up to 50 sequences over 24 hours.

The new MAS-100NT® Ex systems are controlled using a new menu driven, larger illuminated display allowing for quicker navigation. A programmable start delay of up to 60 minutes allows for personnel to be out of the sampling area when the sampling starts and a new audible alarm indicating the interruption of a sampling.

 

Advantages

 

Technical Specification

Feature

Specification

Height

25 cm

Diameter

11 cm

Weight

2.38 kg

Material

Anodized aluminum

Nominal Airflow

100 SLPM ±2.5 % (standard litres per minute)

Pre-set sampling volumes

50, 100, 250, 500 and 1,000 litres

User defined sampling volume range

1 to 2,000 litres, in increments of 1 liter

Battery pack

Li-ion, rechargable battery pack

Charging time

Full recharge time approx. 3.5 hours

Running time

Total running time up to 7 hours

Total aspiration volume

Approx. 42,000 liters

Motor

6 V

Display

Backlit liquid crystal display

Battery

RTC (Real Time Clock) battery; lifetime of approx. 10 years

Driving motor

PWM (Pulse Width Modulation) frequency for driving motor

Airflow regulation

Hot wire anemometer, numerical control, temperature and pressure sensors

CE approval

Emission: EN 61326-1:2006, EN 55011:1998+A1:99
Immunity: EN 61326-1:2006, EN 61000-4-2:1995 + A1:98 +A2:01, EN 61000-4-3:2002, EN 61000-4-4:1995 + A1:01 +A2:01, EN 61000-4-5:1995 + A1:01, EN 61000-4-6:1996 + A1:01, EN 61000-4-8:1993 + A1:0

Power unit / battery charger

110-240 V, 50-60 KHz

Data exchange

USB interface

 

Downloads

 

More detail about Muser Apac Sdn Bhd
Muser Apac Sdn Bhd
Muser Apac Sdn Bhd Testing Instruments Supplier Kuala Lumpur (KL), Malaysia, Monitoring Mapping Systems Supply Selangor, Measuring Equipments Supplies ~ Muser Apac Sdn Bhd
Contact Us flagMalaysia