Function | Range | Resolution | Accuracy | |
AC volts (40 Hz to 500Hz)1 |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
1.0% + 3 | |
DC volts |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
0.5% + 3 | |
AC millivolts | 400.0 mV | 0.1 mV | 3.0% + 3 | |
DC millivolts | 400.0 mV | 0.1 mV | 1.0% + 10 | |
Diode test2 | 2.000 V | 0.001 V | 10% | |
Resistance (Ohms) |
400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ |
0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
0.5% + 3 0.5% + 2 0.5% + 2 0.5% + 2 0.5% + 2 1.5% + 3 |
|
Capacitance3 |
40.00 nF 400.0 nF 4.000 μF 40.00 μF 400.0 μF 1000 μF |
0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF |
2% + 5 2% + 5 5% + 5 5% + 5 5% + 5 5% + 5 |
|
AC current μA (40 Hz to 400 Hz) |
400.0 μA 4000 μA |
0.1 μA 1 μA |
1.5% + 3 | |
AC current mA (40 Hz to 400 Hz) |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5% + 3 | |
AC current A (40 Hz to 400 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5% + 3 | |
DC current μA |
400.0 μA 4000 μA |
0.1 μA 1 μA |
1.5% + 3 | |
DC current mA |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5% + 3 | |
DC current A |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5% + 3 | |
Backlight | Yes | |||
1All ac, Hz, and duty cycle are specified from 1% to 100% of range. Inputs below 1% of range are not specified. 2Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA. 3Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 40 nF range). |
Function | Overload protection | Input impedance (Nominal) | Common mode rejection ratio | Normal mode rejection ratio |
AC volts | 1000 V 1 | >10 MΩ <100 pF |
>60 dB at dc, 50 Hz or 60 Hz |
- |
AC millivolts | 400 mV | >1MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
DC volts | 1000 V 1 | >10 MΩ <100 pF |
>100 dB at dc, 50 Hz or 60 Hz |
>60 dB at 50 Hz or 60 Hz |
DC millivolts | 400 mV | >1MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
1 106 V Hz Max |
General specifications | |
Maximum voltage between any terminal and earth ground | 1000 V |
Display (LCD) | 4000 counts, updates 3/sec |
Battery type | 2 AA, NEDA 15A, IEC LR6 |
Battery life | 500 hours minimum (50 hours in LED Test mode without load. The hours with load depends on the type of LED under test.) |
Temperature | |
Operating | 0 °C to 40 °C |
Storage | -30 °C to 60 °C |
Relative humidity | |
Operating humidity |
Non-condensing (<10°C) ≤90% RH at 10 °C to 30 °C; ≤75% RH at 30 °C to 40 °C |
40 MΩ range ≤80% RH at 10 °C to 30 °C; ≤70% RH at 30 °C to 40 °C |
|
Altitude | |
Operating | 2000 m |
Storage | 12,000 m |
Temperature coefficient | 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
Fuse protection for current inputs |
440 mA, 1000 V Fast Fuse, Fluke specified part only. 11A, 1000V Fast Fuse, Fluke specified part only. |
Size (H x W x L) | 183 x 91 x 49.5 mm |
Weight | 455 g |
IP rating | IP 40 |
Safety | IEC 61010-1, IEC61010-2-030 CAT III 600 V, CAT II 1000 V, Pollution Degree 2 |