TRIOS VIPER

TRIOS VIPER

Category: TRIOS Available
For more information, visit our official website at hsa-asia.com

Specifications
  • VIPER VIS PHOTOMETER FOR MEASUREMENTS OF HYPERSPECTRAL ATTENUATION AND TRANSMISSION: .
Description

VIPER measures hyperspectral attenuation and transmission coefficients in the wavelength range of 360 nm and 750 nm, enabling detailed determination of multiple parameters at the same time. The light source is provided by 5 selected, energy-saving LEDs that Warranty a long service life and stable measurement data. VIPER can be used in different media as it is available in multiple path lengths, both in stainless steel or titanium housing.

Typical applications for VIPER are water quality monitoring, color measurements of aqueous solutions or quality monitoring of drinking water. Like all TriOS sensors, VIPER is equipped with a nano-coated optical window that protects from fouling. Additional parameters can be installed by means of software if necessary at a later time.

Specification

 
Light Source
 
5 LEDs
 
Detector
High-end miniature spectrometer, 256 channels
360 to 750 nm, 2.2 nm/pixel
Measurement principle
Attenuation
Optical path
10 mm, 50 mm, 100 mm, 150 mm, 250 mm
 
 
Parameter
SAC436
Pt-Co color scale (APHA/Hazen) (390 nm, 455 nm)
Colouring based on DIN EN ISO 7887-C (410 nm, 436 nm, 525 nm, 620 nm)
Cr-Co color scale (380 nm, 413 nm)
Measuring range
0.01...2.5 AU (absorption units)
Measurement accuracy
< 0.2 %
Turbidity compensation
Yes
Data logger
~ 2 MB
T100 response time
min. 4 s
Measurement interval
min. 2 s
 
Housing material
Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø)
495 mm x 48 mm (with 50 mm path)
~ 19.5 x 1.9 (with 50 mm path)
Weight stainless steel
~ 2.4 kg (with 50 mm path)
~ 5.3 lbs (with 50 mm path)
Weight titanium
~ 1.3 kg (with 50 mm path)
~ 2.9 lbs (with 50 mm path)
 
Interface digital
Ethernet (TCP/IP), RS-232 or RS-485 (Modbus RTU)
Power consumption
≤ 1 W
Power supply
12...24 VDC (± 10 %)
 
Maintenance effort
≤ 0.5 h/month (typical)
Calibration/maintenance interval
24 months
System compatibility
Modbus RTU
Warranty
1 year (EU: 2 years)
 
US: 2 years
 
Max. pressure with SubConn
30 bar
~ 435 psig
Max. pressure with fixed
3 bar
~ 43.5 psig
Max. pressure in FlowCell
1 bar, 2...4 L/min
~ 14.5 psig, 0.5 to 1.0 gpm
Protection type
IP68
NEMA 6P
 
Probentemperatur
+2...+40 °C
~ +36 F to +104 F
Umgebungstemperatur
+2...+40 °C
~ +36 F to +104 F
Lagertemperatur
-20...+80 °C
~ -4 F to +176 F
Anströmgeschwindigkeit
0,1...10 m/s
~ 0.33 fps to 33 fps


Measuring Range
 
parameter
according to
unit
factor
path length [mm]
 
 
 
 
10
50
100
150
250
SAC436
DIN EN ISO 7887:2011_method B
1/m
-
1...250
0.2...50
0.1...25
0.06...17
0.04...10
SAC525
DIN EN ISO 7887:2011_method B
1/m
-
1...250
0.2...50
0.1...25
0.06...17
0.04...10
SAC620
DIN EN ISO 7887:2011_method B
1/m
-
1...250
0.2...50
0.1...25
0.06...17
0.04...10
True Color 410
DIN EN ISO 7887:2011_method C
mg/L Pt
18.52
20...3750
4...750
2...375
1.2...250
0.8...150
Pt-Co-Color 390
DIN EN ISO 6271-2016:05
mg/L Pt
7.4
8...1500
1.6...300
0.8...150
0.4...100
0.2...60
Pt-Co-Color 455
DIN EN ISO 6271-2016:05
mg/L Pt
36.4
40...7500
8...1500
4...750
2.4...500
1.4...300
Cr-Co-Color 380
-
° (color degree)
9.7
10.0...2000
2...400
1...200
0.6...130
0.4...80
Cr-Co-Color 413
Gost 3351-74
° (color degree)
34.1
40...7000
8...1400
4...700
2.6...450
1.6...275


Versions

More detail about HSA ASIA SDN BHD
HSA ASIA SDN BHD
HSA ASIA SDN BHD Measuring Instruments Selangor | Flow Meter Supply Kuala Lumpur (KL) | Water Analytical Instruments Supplier Malaysia ~ HSA ASIA SDN BHD
Contact Us flagMalaysia