This 3-part hematology analyzer is used for detecting the parameters of RBC, WBC, HGB and differential counting.
Features:
Models:
Model |
BK-6190 |
||||
Throughput |
60Tests/hour |
||||
Measuring Principle |
Impedance for cell counting Cyanide-free method for HGB |
||||
Aperture Diameter |
WBC chamber:100μm,RBC/PLT chamber:70μm |
||||
Parameters |
WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR |
||||
Histograms |
3 histograms(includingWBC, RBC, PLT histogram) |
||||
Sample Volume |
Normal:10μl; Predileted:20μl |
||||
Calibration |
Manual & auto calibration |
||||
Counting Modes |
Venous, capilary, prediluted |
||||
Printer |
Built-in thermal printer,USB to external printer(optional) |
||||
Data Input |
8 inch touch screen, keyboard, mouse |
||||
Output |
RS232 port, USB port, VGA port. LIS system avaliable |
||||
Storage |
100,000 results with histogram |
||||
Carry-over rate |
Parameter |
CV |
|||
WBC |
≤ 2% |
||||
RBC |
≤ 1% |
||||
HGB |
≤ 2% |
||||
PLT |
≤ 2% |
||||
Background |
Parameter |
Background |
|||
WBC |
≤ 0.2×10^9/L |
||||
RBC |
≤ 0.02×10^12/L |
||||
HGB |
≤ 1 g/L |
||||
HCT |
≤ 0.5% |
||||
PLT |
≤ 10×10^9/L |
||||
Precision
|
Parameter |
CV |
Measurement range |
||
WBC |
≤ 2.0% |
(4.0-15.0×10^9)/L |
|||
RBC |
≤ 1.9% |
(3.5-6.0×10^12)/L |
|||
HGB |
≤ 1.9% |
(110.0-180.0g)/L |
|||
MCV |
≤ 0.4% |
(80.0-110.0)fL |
|||
PLT |
≤ 4.0% |
(100.0-500.0×10^9)/L |
|||
Linearity |
Parameter |
Measurement range |
CV |
||
WBC |
(0-5.9×10^9)/L (6.0-99.9×10^9)/L |
(±0.3×10^9)/L ±5% |
|||
RBC |
(0-0.99×10^12)/L (1.0-9.99×10^12)/L |
(±0.05×10^12)/L ±5% |
|||
HGB |
(0-99.0g)/L (100.0-300.0g)/L |
(±2.0g)/L ±2% |
|||
PLT |
(0-99×10^9)/L 100-999×10^9)/L |
(±8.0×10^9)/L ±10% |
|||
Power supply |
AC 220±10%,60/50Hz, 110±10%,60Hz |
||||
Package Size |
580*470*590mm (W*D*H) |
ReagentPackage Size |
450*410*310mm (W*D*H) |
||
Gross Weight |
27kg |
ReagentGross Weight |
23kg |