Description
Features
Simple User Interface
Operation is simple with limited features that only require the use of a couple of buttons. Readings are easy to view on the large, clear display.
One or Two-Point Calibration
Automatic calibration can be performed at 1 or 2 points using the 5 preprogrammed buffers (4.01, 6.86, 7.01, 9.18, and 10.01).
HI1131B pH Electrode
The HI2211 is supplied with the HI1131B glass body, double junction, refillable pH electrode with an indicating sensor made of High Temperature (HT) glass. The double junction and HT glass design allows the HI1131B to be used in a wide variety of application from samples with metals and Tris buffer to samples at elevated temperatures.
Temperature Compensation
The HI2211 offers automatic or manual temperature compensation of pH measurements over a wide range from -9.9 to 120.0°C.
mV mode
The HI2211 has a mV mode that can be used with a wide variety of ORP electrodes. The mV mode is also useful in checking the offset and slope of the pH electrode. Understanding the offset and slope characteristic is helpful in knowing if the probe is functioning correctly.
Stability Indicator
The HI2211 features a stability indicator. An hourglass appears on the display when there is instability in the reading. The hourglass disappears once the reading has stabilized. At that time a reading should be taken or stored.
Measurement Recall
The HI2211 has a “MEM” button to store a reading. The reading can be recalled at anytime by pressing the memory recall “MR” button.
Specification
Specification Name |
Detail |
SKU |
HI2211-01 |
pH Range pH |
-2.00 to 16.00; ±399.9 mV ; ±2000 mV |
pH Resolution |
0.01 pH; 0.1 mV; 1 mV |
pH Accuracy |
(@25°C/77°F) ±0.01 pH; ±0.2 mV (±399.9 mV); ±1 mV (±2000 mV) |
pH Calibration |
automatic, one or two-point with five memorized buffer values (pH 4.01, 6.86, 7.01, 9.18, 10.01) |
pH Temperature |
Compensation automatic or manual from -9.9 to 120.0°C |
Temperature Range |
-9.9 to 120.0°C (14.2 to 248.0°F) |
Temperature Resolution |
0.1 °C |
Temperature Accuracy |
±0.5 °C (0.0 - 100.0 °C); ±1 °C (outside) |
pH Electrode |
HI1131B pH electrode |
Temperature Probe |
HI7662 temperature probe |
Power Supply |
12 VDC adapter (included) |
Environment |
1 to 50°C (32 to 122°F), RH max 95% non-condensing |
Dimensions |
240 x 182 x 74mm (9.4 x 7.1 x 2.9”) |
Weight |
1.3 Kg (2.9lbs) |
Ordering Information |
HI2211 is supplied with HI1131B pH electrode,
HI7662 temperature probe, HI76404N electrode holder,
HI70004 pH 4.01 buffer solution sachet,
HI70007 pH 7.01 buffer solution sachet,
HI7082 electrolyte solution,
HI700601 cleaning solution sachet, 12 VDC adapter, and instructions |
Warranty |
2 years
|
More detail about ENVIROLAB SDN. BHD.