Specification |
Benefit |
Detector |
|
G4 PHOENIX |
Thermal conductivity detector with reference channel and adjustable gain amplifier |
Reliable, adjustable ranges, drift free |
G4 PHOENIX MS |
Mass spectrometer, m/z range 1 - 100 amu, single quadrupole, optimized EI source and channeltron detector |
Unique mass spectrometry performance |
Furnaces |
|
Infrared Heated |
IR furnace up to 900 ºC, i.d. of quartz tube 30 mm, water cooled, optinal thermocouple kit for direct sample temperature readings |
Precise temperature control, flexible heating programms, accepts large samples |
Resistance Heated (Option) |
Additional resistance heated furnace up to 1100 ºC, i.d. of quartz tube 18 mm |
Residual hydrogen in multi-phase duplex steels |
Carrier Gas |
Nitrogen 99.995% purity, min. 2 bar (~ 50 psi), 99.9990% purity for trace analysis |
Pre-cleaning using regenerateable molecular sieve |
Calibration Gas |
With pure gases (H2 or He) or certified mixtures (each component 99.999% purity), automatic gas dose calibration system with 10 individual volumes |
Easy, precise gas calibration without standards, traceable to P, T, V |
Cooling Water |
1 l/min @ 3 bar (44 psi) |
Fast cool down, standard tap water compatible, water-saving design with water-stop valve, chiller also possible |
Power Supply |
|
G4 PHOENIX |
230 VAC (± 10 %), 50-60 Hz, 2200 VA |
Industry standard power and current configurations |
Mass Spectrometer |
230 VAC, 50-60 Hz, 250 VA |
|
Dimension & Weight |
|
G4 PHOENIX |
630 x 700 x 670 mm (W x D x H), weight ~ 50 kg |
|
Mass Spectrometer |
630 x 640 x 480 mm (W x D x H), weight ~ 60 kg |
|