HIOKI IM7581 Impedance Analyzer

HIOKI IM7581 Impedance Analyzer

Category: Impedance Analyzers Available
For more information, visit our official website at marktest.com.my

Description

HIOKI IM7581 Impedance Analyzer
100 kHz to 300 MHz Measurement Frequency Impedance Analyzer with 0.5 ms Test Speed

Hioki LCR Meters and Impedance Analyzers range from 1 mHz to 3 GHz devices to suit a wide range of applications in the testing of electronic components. The IM7581 offers a top measurement time of 0.5 ms over a 100 kHz to 300 MHz frequency range, making it ideal for high volume production of ferrite chip beads and chip inductors.

LCR Measurement Using the SMD Test Fixture IM9201: Impedance Analyzer IM7580 series
The IM7580 series encompasses five models covering measurement frequencies of 1 MHz to 3 GHz. Used in combination with the Test Fixture IM9201, which can accommodate six SMD sizes, IM7580 series instruments allow you to measure specimens easily and reliably.

Key Features
The instrument does not ship with a test fixture or probe. A test fixture designed specifically for use with the Impedance Analyzer is required.


FEATURES
 
Contact check function: Impedance Analyzer IM7580 series
The instruments’ contact check function lets you check contact between the specimen and measurement terminals to detect faulty contact or verify good contact.

Comparator function: Impedance Analyzer IM7580 series
A comparator function lets you check whether measured values fall within a user-defined judgment range. This functionality is ideal for use in generating PASS/FAIL judgments for specimens.

How to connect the test head: Impedance Analyzer IM7580 series
Connect the cable of the test head to the impedance analyzer.
Tighten the nut until the handle of the torque wrench bends slightly.
Do not over tighten.

Application information
Download the sample application and view the Communications Command User Manual.

Click here for the IMPEDANCE ANALYZER application.

Accuracy calculation information
Calculate accuracy by simply entering a numerical value.

Click here to calculate accuracy.


SPECIFICATIONS

Basic specifications
Accuracy guaranteed: 1 year
 
Measurement modes LCR mode, Analyzer mode (Sweeps with measurement frequency and measurement level), Continuous measurement mode
Measurement parameters Z, Y, θ, Rs (ESR), Rp, X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q
Measurable range 100 mΩ to 5 kΩ
Display range Z: 0.00 m to 9.99999 GΩ / Rs, Rp, X: ± (0.00 m to 9.99999 GΩ)
Ls, Lp: ± (0.00000 n to 9.99999 GH) / Q: ± (0.00 to 9999.99)
θ: ± (0.000° to 180.000°), Cs, Cp: ± (0.00000 p to 9.99999 GF)
D: ± (0.00000 to 9.99999), Y: (0.000 n to 9.99999 GS)
G, B: ± (0.000 n to 9.99999 GS), Δ%: ± (0.000 % to 999.999 %)
Basic accuracy Z: ±0.72 % rdg. θ: ±0.41°
Measurement frequency 100.00 kHz to 300.00 MHz (5 digits resolution)
Measurement signal level Power: -40.0 dBm to +7.0 dBm
Voltage: 4 mV to 1001 mVrms
Current: 0.09 mA to 20.02 mArms
User-configured power, voltage, and current
Output impedance 50 Ω
Display 8.4-inch color TFT with touch screen
Measurement speeds *1 FAST: 0.5 ms / MED: 0.9 ms / SLOW: 2.1 ms / SLOW2: 3.7 ms
*1 Analog measurement time
Functions Contact check, Comparator, BIN measurement (classification), Panel loading/saving, Memory function, Equivalent circuit analysis, Correlation compensation
Interfaces Handler, USB, LAN, GP-IB (optional), RS-232C (optional)
Power supply 100 to 240 V AC, 50/60 Hz, 70 VA max.
Dimensions and mass Main unit: 215 mm (8.46 in) W × 200 mm (7.87 in) H × 268 mm (10.55 in) D, 6.5 kg (229.3 oz)
Test head: 61 mm (2.40 in) W × 55 mm (2.17 in) H × 24 mm (0.94 in) D, 175 g (6.2 oz)
Included accessories Test head ×1, Connection cable ×1, Power cord ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1

More detail about Marktest Technology Sdn Bhd
Marktest Technology Sdn Bhd
Marktest Technology Sdn Bhd Measurement Equipment Supplier Johor, Laser Marking System Malaysia, Power Quality Analyzer Supply ~ Marktest Technology Sdn Bhd
Contact Us flagMalaysia