Genie A Ultrapure & RO Lab Water Systems is a plug-and-play lab water system with a small footprint, producing ultrapure water (Type I water) and RO water from tap water directly. It is an ideal choice for users who need up to 20 liters of ultrapure water or RO water per day. Quality of ultrapure product water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards
The system is CE/RoHS certified. RephiLe provides full documents support to meet user’s GMP, GSP, GAP, GCP, GLP requirements.
Design & Functions
Effective TOC level controlment
Easy Operation and Maintenance
Cartridges and Parts
1. Cartridge
Pre-filtration pack removes most large particles, chlorine, calcium or organic compounds to protect RO membrane to extend their usable life.
Ultra purification cartridge is filled with LeFilTM and OrganeFilTM media to remove trace ions and organics.
2. Control console
8-inch console controls system and peripheral devices (Such as leak protector and tank sanitization module). All operations can be done on the console by touching the screen.
The control console can be placed on the bench or inside a drawer for further flexibility and space saving.
The console screen and dispenser screens are water-drop proof. You can operate the console and dispenser with latex gloves on.
The system has two level password protection on display menu for safety purposes.
3. Dispenser
The volumetric function can automatically deliver the water volume you need once it is set up.
With the adjusting button on a control console or a dispenser, you can dispense water at the flow rate you need up to 2 liters per minute.
A dispenser handle can be set on the dispenser stand, on the main system, or even hanging onto other places to free up maximum bench space.
4. Final Filter
A wide range of final filters (optional) is applicable to ensure ultrapure water without particles, bacteria or pyrogen.
Ultrapure Water
RO Pure Water
Feed Water Requirements | |
Tap water | Municipal water, Conductivity < 1000 μS/cm (500 ppm) |
Operating temperature | 5 - 35 ℃ |
Feed water pressure | 1 - 6 bar (15 - 90 psi) |
Water Flow Rate | |
RO water production rate (@ 25℃) | 32 L/hr |
RO water dispensing rate | 0.6 L/min |
Ultrapure water dispensing rate | 0.6 L/min |
Product Water Quality | |
RO water conductivity (@ 25℃) | typically < 20 μS/cm (feed water conductivity < 500 μS/cm) |
RO rejection | 97 to 98% ionic rejection (new RO cartridge) > 99% organic rejection > 99% particulates & bacteria rejection |
Ultrapure water resistivity (@ 25℃) | 18.2 MΩ·cm |
Ultrapure water TOC* | < 2 ppb |
Particles in ultrapure water ( > 0.2 μm) | No particles (with a 0.2 μm final filter or RephiBio filter) |
Microorganisms in ultrapure water | < 0.01 cfu/ml (with a 0.2 μm final filter or RephiBio filter) |
Pyrogens (endotoxins) in ultrapure water | < 0.001 Eu/ml (with a RephiBio filter) |
RNase in ultrapure water | < 0.5 pg/ml (with a RephiBio filter) |
DNase in ultrapure water | <10 pg/ml (with a RephiBio filter) |
Dimensions | |
Main system dimensions Width × depth × height | 32 × 44 × 54 cm (12.6 × 17.3 × 21.3 in) |
Dispenser dimensions Width × depth × height | 21 × 29 × 61 cm (8.3 × 11.4 × 24.0 in) |
Weight | 19 kg |
Input Voltage | 110 - 240 VAC |
Operating Voltage | 24 VDC |
Main system power | < 200 W |
*In the appropriate operating conditions, otherwise typically ≤5 ppb.