InKjel Programmable Infrared Digestion Device InKjel 450 P, InKjel 475 P, InKjel 625 P, InKjel 1210 P, InKjel 1225 P

InKjel Programmable Infrared Digestion Device InKjel 450 P, InKjel 475 P, InKjel 625 P, InKjel 1210 P, InKjel 1225 P

Category: Nitrogen Determination according to Kjeldahl Available
For more information, visit our official website at sciencegates.com

Description

InKjel Programmable Infrared Digestion Device InKjel 450 P, InKjel 475 P, InKjel 625 P, InKjel 1210 P, InKjel 1225 P

Infrared rapid digestion devices for nitrogen determination according to Kjeldahl

The digestion units of the model series behrotest® InKjel are equipped with a high-performance infrared heater. Quality and position of the behr infrared heater guarantee the user identical heating phases and digestion temperatures at all sample positions.

This also applies to the double-row arrangement for racks for 12 samples. The glass digestion vessels are suspended in the rack and are not deposited at the bottom of the digestion unit. Therefore the digestion units are less susceptible to breakage than in an aluminum heating block.

This makes the behrotest® InKjel the ideal digestion system for nitrogen determination according to Kjeldahl and other high-temperature digestions. In the Kjeldahl digestions, the temperature is adjusted through the boiling point of sulfuric acid.


Technical data
 
 
InKjel 450 P
InKjel 475 P
InKjel 625 P
InKjel 1210 P
InKjel 1225 P
Item number
B00218107
B00218108
B00218105
B00497540
B00218106
Packaging quantity
1
Width (in cm)
54.00
Depth (in cm)
44.00
Height (in cm)
75.00
Weight (in kg)
20.00
Power consumption (in W)
1500
Current consumption (in A)
8
Sample positions
4
4
6 12 12
Digestion vessel size (in ml)
500
750 250 100 250
Energy setting range
0 - 100 %, infinitely variable, manually adjustable
Time setting range
0 - 199 min, in steps of 1 min.


Benefits

Digestion with infrared
 
  1. Homogeneous heating
  2. The special benefit of our behrotest Inkjels is that our infrared heaters are heated evenly along the entire heating route.
  3. Infrared digestions compared with block digestion
  4. Particularly homogeneous heating of samples with lateral heaters, so there are no zones with different temperatures. Hardly any boiling delays.
  5. Heating of samples from below, therefore higher requirements for the user for avoiding boiling delays.
  6. Quartz heater:
    - Uniform effect along the entire length
    - Isolated from casing, no loss of energy as a result of heat transfer
    - Full heating power within 1 minute, no preheating required
 

Areas of use
 
Food analysis › Beverage industry › Feed analysis › Agricultural industry › Environmental analysis › Pharmaceuticals industry

Parameter to be analyzed
 
Phosphorous

Sample material
 
Feed › Sludges › Water


Ordering Information
Model
Description
Part Number
InKjel 450 P
Programmable infrared digestion system for 4 glass vessels of 500 ml
B00218107
InKjel 475 P
Programmable infrared digestion system for 4 glass vessels of 750 ml
B00218108
InKjel 625 P
Programmable infrared digestion system for 6 glass vessels of 250 ml
B00218105
InKjel 1210 P
Programmable infrared digestion system for 12 glass vessels of 100 ml
B00497540
InKjel 1225 P
Programmable infrared digestion system for 12 glass vessels of 250 ml
B00218106